S4a XDUCER TYPE:
1. THC-HS
2.THC-RS2
STEP5.
INSTALL XDUCER NOW
Dist= 113.41mm, V
Sound Speed in Pipe Wall
: If your pipe material is on the list in the previous step, the flow meter will
automatically skip this step and go to the next step. You do not have to enter in the Sound Speed
unless you have selected Other as your pipe material.
If you selected Other as your pipe material, you will have to enter the sound speed manually. Press
ENT which will prompt you to enter the correct sound speed for the pipe wall material you are using.
You can find this data in the Appendix §9.3 of the User’s Manual. When you are done, press ENT to
confirm.
Pipe lining
: Press ▼/- to scroll down, and then, press ENT. If your pipe
has lining inside, enter the lining information. Press ENT to confirm. Press
(“Ex”) to go back to the Quick Start menu.
Step 3: Enter fluid info
From the Quick Start menu, scroll down to Step 3.
Fluid Type
: Press ENT and select the item that matches your fluid type. If
you do not find a match (non-standard fluid), select item 9 (Other). Press
the ENT key to confirm.
Sound Speed in Fluid
: If you found your fluid type in the previous step, the flow meter already has
the sound speed data. Therefore, skip this step and go on to the next.
Otherwise, exit to the main screen and press keys S32. Press ENT and key in the sound speed of your
fluid. You can find this information in Appendix §9.3 of the User’s Manual. When you are done,
press ENT to confirm.
Step 4: Enter transducer installation info
From the Quick Start menu, scroll down to Step 4.
Transducer Type
: Press ENT to edit the type of transducer. Use the ▼/- key to scroll down and
press ENT and select the proper transducer type from the list.
Note: The type of transducer you are
using will be printed on the side of the transducers.
Press ENT to confirm.
Mounting Method
: Use the ▼/- key to scroll down to the mounting method
selection screen. Then, press ENT and select the proper method. Press ENT
to confirm. For pipes smaller than 1”, try the W-method. For pipes from 1” to
12”, use the V-method. For pipes larger than 12”, use the Z-method. See
Section §3.3 for more details on how to select the proper method.
Mounting Spacing
: Use the ▼/- key to scroll down to Step 5. The displayed
value is the mounting spacing between the two transducers (see the image on
the right).
Write down this number, as you will need it later when
installing the transducers.
S2d: Has Liner?
3. Rubber
S3a
Select Fluid
9. Other
1.StainlessSteel
Содержание RH40
Страница 62: ...62...
Страница 63: ...63...
Страница 64: ...64...
Страница 65: ...65...
Страница 66: ...66...
Страница 67: ...67...
Страница 68: ...68...
Страница 69: ...69...
Страница 70: ...70...
Страница 71: ...71...
Страница 72: ...72...
Страница 73: ...73...
Страница 74: ...74 8 4 Sound Speed Tables...
Страница 75: ...75...
Страница 76: ...76...
Страница 77: ...77...
Страница 78: ...78...
Страница 79: ...79...
Страница 80: ...80...
Страница 81: ...81...
Страница 82: ...82...
Страница 83: ...83...
Страница 84: ...84...
Страница 85: ...85...
Страница 86: ...86...
Страница 87: ...87...
Страница 88: ...88...
Страница 89: ...89...
Страница 90: ...90...