RCL40 & RCL41
Installation manual
Regasense is a part of Regal Components |
Rev. no.
2021_1
9
Installation of control box
General information
Both RCL40 and RCL41 sensors can be used with both RCL41 and RCL43 control boxes. Choose the control
box best suited for your application. RCL41 control box is designed for DIN mounting and RCL43 control box
is designed with a sealed aluminium housing. The controller can be placed up to 50 meters away from the
sensor.
Do not supply power to the controller until all wires or cables are connected properly for selected
output.
Connect the wires from the sensor, power supply and 0V as well as choosen output configuration
before the power is switched on. Find the schematic under the instructions for each controller.
Always use shielded cables to connect the sensor probe and keep in mind that the cable shield on
RCL41 and RCL41P sensor is connected to the header of the sensor. That may cause damage if there is a
potential difference between the location of the sensor and the grounding point of the cable shield.
To ensure proper EMC-performance:
- The shield of the sensor should be connected to an EMC-cable gland when entering the cabinet in which
the controller is placed.
- The sensor wires inside the cabinet should be kept as short as possible.
- If the cabinet contains EMC-noisy equipment there may be need to EMC-protect the wires and the con-
trol box.