reedinstruments
www
com
Peak Hold button
The.Peak.Hold.function.captures.the.peak.AC.or.DC.voltage.or.current..
The.meter.can.capture.negative.or.positive.peaks.as.fast.as.1.millisecond.
in.duration..
1..Turn.the.function.switch.to.the.A.or.V.position.
2..Use.the.
MoDE
.button.to.select.AC.or.DC.
3..Allow.time.for.the.display.to.stabilize.
4..Press.and.Hold.the.
PEaK
.button.until.“CAL”.appears.in.the.display..
This.procedure.will.zero.the.range.selected.
5..Press.the.
PEaK
.button,.Pmax.will.display.
6..The.display.will.update.each.time.a.higher.positive.peak.occurs.
7..Press.the.
PEaK
.button.again,.Pmin.will.display..The.display.will.now.
update.and.indicate.the.lowest.negative.peak.
8..To.return.to.normal.operation,.press.and.hold.the.
PEaK
.button.until..
the.Pmin.or.Pmax.indicator.switches.off.
noTE:
.If.the.Function.switch.position.is.changed.after.a.calibration.the.
Peak.Hold.calibration.must.be.repeated.for.the.new.function.
selected.
RANGE button
When the meter is first turned on, it automatically goes into AutoRanging.
This.automatically.selects.the.best.range.for.the.measurements.being.
made.and.is.generally.the.best.mode.for.most.measurements...
For.measurement.situations.requiring.that.a.range.be.manually..
selected,.perform.the.following:
1..Press.the.
RanGE
.button..The.“Auto.Range”.display.indicator.will..
turn.off,.The.“Manual.Range”.display.indicator.will.turn.on
2..Press.the.
RanGE
.button.to.step.through.the.available.ranges.until..
you.select.the.range.you.want.
3..Press.and.hold.the.
RanGE
.button.for.2.seconds.to.exit.the..
ManualRanging.mode.and.return.to.AutoRanging.