background image

reedinstruments

www

com

8

Cleaning

•  Periodically wipe the case with a dry cloth. Don’t use abrasives 

 

or solvents on this instrument

Notes   

How it Works

This REED Infrared Thermometer measures the surface temperature of an 
object. The unit’s optics sense emitted, reflected, and transmitted energy 
which is collected and focused onto a detector. The unit’s electronics 
translate the information into a temperature reading which is displayed on 
the unit. In units with a laser, the laser is used for aiming purposes only.

Field of View

Make sure that the target is larger than the unit’s spot size. The smaller 
the target, the closer you should be to it. When accuracy is critical, make 
sure the target is at least twice as large as the spot size.

Distance & Spot Size 

As the distance (D) from the object increases, the spot size (S) of the area 
measured by the unit becomes larger. See Graphic on Page 4.

Locating a Hot Spot

To find a hot spot aim the thermometer outside the area of interest, then 
scan across with an up and down motion until you locate hot spot.

Reminders

1. Not recommended for use in measuring shiny or polished metal

 

  surfaces (stainless steel, aluminum, etc.). See Emissivity (page 9)
2. The unit cannot measure through transparent surfaces such as

 

  glass. It will measure the surface temperature of the glass instead
3. Steam, dust, smoke, etc., can prevent accurate measurement 

 

  by obstructing the unit’s optics

continued ...

Содержание ST-8839

Страница 1: ...reedinstruments www com Model ST 8839 Non Contact Infrared Thermometer Instruction Manual...

Страница 2: ...Function 7 Switching Settings 7 Cleaning 8 Notes 8 9 Battery Replacement 10 Safety Use extreme caution when the laser beam is turned on Do not let the beam enter your eye another person s eye or the...

Страница 3: ...cifications Response Time Less than 1 second Spectral Response 8 14um Emissivity Digitally adjustable from 0 10 to 1 0 Over Range Indication LCD will show 0L 0L Polarity Automatic no indication for po...

Страница 4: ...ase instruction manual Notes Accuracy Given at 18 to 28 C 64 to 82 F less than 80 RH Field of View Make sure that the target is larger than the unit s spot size The smaller the target the closer you s...

Страница 5: ...on 8 Measurement Trigger 9 Handle Grip 10 Battery Cover Display Description 1 Data Hold 2 Measuring indication 3 Emissivity symbol value 4 Low battery lock laser on 5 Backlight on symbol 6 C F symbol...

Страница 6: ...begin testing The display will light if the battery is good Replace the battery if the display does not light 3 While measuring the SCAN display icon will appear in the upper left corner of the LCD 4...

Страница 7: ...ect the temperature units C or F by using the C F switch To lock the unit on for continuous measurement slide the middle switch LOCK ON OFF to the right If the trigger is pulled while the unit is lock...

Страница 8: ...e target the closer you should be to it When accuracy is critical make sure the target is at least twice as large as the spot size Distance Spot Size As the distance D from the object increases the sp...

Страница 9: ...erature as the material underneath it Measure the temperature of the tape or painted surface Emissivity Values Substance Thermal Emissivity Substance Thermal Emissivity Asphalt 0 90 to 0 98 Cloth blac...

Страница 10: ...o reedinstruments com Notes __________________________________________ ________________________________________________ ________________________________________________ _______________________________...

Страница 11: ..._______________ ________________________________________________ ________________________________________________ ________________________________________________ _____________________________________...

Страница 12: ..._______________ ________________________________________________ ________________________________________________ ________________________________________________ _____________________________________...

Отзывы: