The current alarm value will appear on the LCD. Press the “UP” and 
“DOWN” buttons to modify the ppm value by increments of 100. The 
alarm ranges from 100 to 9900ppm. When the desired alarm value is 
displayed, press the “ENTER” button to save, or press the “ESC” button 

to cancel.

When measured CO

2

 values are above the set limit the meter will start 

to beep. Press any button (except for the “POWER” button) to stop the 
beeping, or move the meter to a location where the ppm value is below 
the alarm limit. If the beeping does not stop, restart the meter.

Temperature Unit

To change the temperature scale from °C to °F, press and hold the “SET” 
button for 2 seconds. Press the “DOWN” button and “Unit” and P3.0 will 
appear on the LCD. Press the “ENTER” button, and modify the tempera

-

ture unit by pressing the “UP” or “DOWN” buttons. Press the “ENTER” 
button to save the setting, or press the “ESC” button to cancel. 

CO

2

 Calibration

This meter is calibrated at a standard 400ppm CO

2

 concentration. It’s sug-

gested to perform manual calibration regularly to maintain good accuracy. 
For a more thorough calibration and for maintenance for this product, 
please contact REED Instruments at [email protected].

Do not calibrate this meter in an environment with unknown CO

2

 con-

centrations. Manual calibration should be done in fresh outdoor air that 
is well ventilated and in sunny weather. Turn on the meter and hold down 
the “CAL” and “MODE” buttons simultaneously. “400ppm” and “CAL” will 
blink on the LCD indicating you have activated manual calibration, and will 
blink for approximately 5 minutes. When the blinking stops, CO

2

 calibra-

tion is complete, and the meter will automatically switch back to normal-
measuring mode. To abort the calibration, turn off the meter at any time. 

 

 

NOTE

: Ensure the batteries are with full voltage during the calibration to 

prevent from interruption or failed calibration. 

www.

.com

[email protected]

1.800.561.8187

Содержание R9900

Страница 1: ...reedinstruments www com Model Instruction Manual Indoor Air Quality Meter R9900 Rich Black 20 20 20 100 Only if you REALLY need them www com information itm com 1 800 561 8187...

Страница 2: ...w 0 27 100 0 Blue 100 80 30 5 Only if you REALLY need them Table of Contents Features 3 Specifications 3 4 Instrument Description 4 5 Operating Instructions 5 8 Data Hold 5 Backlight 5 Max Min Weighte...

Страница 3: ...minute period and TWA readings Time Weighted Average over a period of 8 hours Hold function freezes displayed reading Maximum minimum and average functions Low drift non dispersive infrared CO2 senso...

Страница 4: ...ne batteries Battery Life 24 hours Includes 4 AA batteries and a hard carrying case Optional accessories 33 Calibration Salt RHA 33 75 calibration salt RHA 75 Instrument Description Power Set button C...

Страница 5: ...powering on the meter will beep and commence a 30 second warm up The meter will then enter the normal measuring mode displaying the current CO2 temperature and humidity readings Readings will be updat...

Страница 6: ...l continue to measure normally NOTE If the meter is turned on for less than 15 minutes the STEL value will be the weighted average of readings taken since power on This is the same for TWA values befo...

Страница 7: ...utton and modify the tempera ture unit by pressing the UP or DOWN buttons Press the ENTER button to save the setting or press the ESC button to cancel CO2 Calibration This meter is calibrated at a sta...

Страница 8: ...the calibrating value of 33 32 7 if at 25 C will blink on the LCD along with the current temperature at the left The meter is now calibrating and will finish when CAL and humidity stop blinking which...

Страница 9: ...ssary Fixed readings Check if the data hold function is activated Slow response Check to see if the airflow channels on the rear are blocked Error messages E01 CO2 sensor damaged E02 The value is unde...

Страница 10: ...n the battery compartment on the rear and make sure they are correct polarity and in good contact Notes __________________________________________ ________________________________________________ ____...

Отзывы: