Maintenance

1. 

If the ambient temperature is not within the automatic temperature  
compensation range (10 - 30°C) the instrument must be calibrated before 
testing commences.

2. 

The prism must be cleaned thoroughly with a soft, clean damp cloth after 
each use - failure to do so will result in inaccurate readings and possible 
long term  
damage.

3. 

The instrument must be kept in a dry clean environment to prevent  
condensation forming on the optical components.

4. 

If used in accordance with the instructions and all warning are followed - 
the optical performance will remain constant and the instrument will give 
many years of reliable service.

Warning

1.  DO NOT IMMERSE IN OR CLEAN WITH RUNNING WATER 
2.  THIS TOOL IS A PRECISION OPTICAL INSTRUMENT, IT REQUIRES  

CAREFUL HANDLING and STORAGE

3.  DO NOT DROP OR SUBJECT TO SHOCK AS DAMAGE MAY RESULT TO 

THE MIRROR TUBE and OPTICAL COMPONENTS

4.  DO NOT SCRATCH OR ATTEMPT TO MEASURE CORROSIVE CHEMICALS 

AS DAMAGE MAY OCCUR TO THE PRISM’S COATING  