9
Auto Power-Off
As a default the instrument will automatically turn off after 15 minutes of
inactivity. To turn off this feature:
1. Turn the meter off.
2. Press the POWER button (to display the full screen).
3. Quickly press the
RANG
button when the instrument is in the
maintenance state as indicated by "1.8.8.888".
4. "AP-XX" will now appear on the display.
5. Press the button to toggle between "AP-ON" & "AP-OF". "AP-OF"
indicates that the no automatic power-off function is disabled, while
"AP-ON" indicates the automatic power-off function is enabled.
6. Press the
ZERO
button to save the required setting.
7. Press and hold the POWER button to exit the maintenance state and
turn the instrument off.
Output Function
• The output terminal of the instrument can produce DC voltages set by
the user or can simulate resistance.
Do not apply any voltage to the output terminal during the operation.
If any improper voltage is applied to the output terminal, it will cause
damage to the internal circuit.
Simulating Output from DC Voltage
1. Insert one end of the test lead into
the output (TC/mV) jack of the
meter and connect the other end
to the input of the instrument under
test, as shown in the diagram to
the right.
2. Press the
FUNC
button to select
"DC Voltage Output". "V" will
appear on the display.
continued...