background image

16

Calibration Procedure

In order to calibrate the R8070SD, an external calibrator that can provide 

a 94.0dB signal is required (REED R8090). If the meter has been exposed 

to extreme conditions or has not been used in a long period of time the 

meter should be calibrated by following steps 1 through 7 below.
1.  Remove the windshield ball.
2.  Power on the meter.
3.  Select "50 - 100dB" range.
4.  Select "FAST" time weighting.
5.  Select "A" weighting.
6.  Insert Microphone into the sound calibrator's microphone adapter.
7.  Power on the sound calibrator (R8090).
If the measured value is not within "94 ±0.2" dB, adjust the meter's 

calibration potentiometer until the display matches the output of the  

calibrator (94.0dB).

Data Stream Output

Using Terminal application, a data stream can be viewed from the RS-232 

output, Connect the RS-232/USB cable between the product and 

terminal/PC and use the following setting to view the data stream.
Terminal Settings:
• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1

41170100000715     =    

+71.5 dB

Reading 
Decimal Place 0-3 

Positive (+) = 0, Negative (-) = 1

dB = 17 
1   

continued...

 

REED Instruments

1-877-849-2127 | [email protected] | www.reedinstruments.com

Содержание R8070SD

Страница 1: ...Data Logging Sound Level Meter Instruction Manual R8070SD REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Страница 2: ...Basic Operation 9 10 Data Hold 10 Peak Hold 10 Recording Maximum and Minimum Readings 10 Backlight 11 Setup Mode 11 13 Setting the Time and Date dAtE 11 Setting Data Decimal Format dEC 12 Enabling Dis...

Страница 3: ...u for purchasing your REED R8070SD Data Logging Sound Level Meter Please read the following instructions carefully before using your instrument By following the steps outlined in this manual your mete...

Страница 4: ...C frequency weighting Fast Slow time weighting Real time data logger with integrated SD memory card User selectable sampling rate from 1 to 3600 seconds Easy to read backlit LCD display Peak hold Dat...

Страница 5: ...low 125ms and 1s Microphone 0 5 12 7mm electret condenser Display 4 digit LCD Backlit Display Yes Data Hold Yes Min Yes Max Yes Alarm Indicators Under and Over visual on screen Datalogging Capabilitie...

Страница 6: ...1 8 245 x 68 x 45mm Weight 0 7lbs 320g Comparison of Frequency Weightings and Tolerance Limits Frequency Hz A Weighting Characteristics C Weighting Characteristics Tolerance IEC 61672 class 2 31 5 39...

Страница 7: ...isplay 3 POWER ESC Backlight Button 4 Up Range Time Check Button 5 Set Fast Slow Button 6 SD Card Slot 7 Hold Next Button 8 REC Enter Button 9 Down A C Sampling Check Button 10 Logger Peak Hold Button...

Страница 8: ...Indicator 3 Data Hold Indicator 4 Record Mode Indicator 5 Maximum and Minimum Indicator 6 Data logging Status Indicator 7 Frequency Weighting Indicator 8 Reading Indicator 9 Date Time Values 10 Unit...

Страница 9: ...vel select SLOW 1 second 3 Press the A C button to select the Frequency weighting Select A Weighting for a general noise sound level or C Weighting for measuring high level noise C Weighting is a stan...

Страница 10: ...utton to capture and freeze a noise peak 2 While in this mode a PEAK symbol will appear and the displayed reading will only change when a higher sound peak is registered 3 Press the LOGGER button agai...

Страница 11: ...iated instructions below Note The meter automatically exits out of the Setup mode if no key is pressed within 7 seconds Setting the Time and Date dAtE 1 Press the ENTER button when dAtE appears on the...

Страница 12: ...activity to preserve battery life 2 Press the ENTER button to confirm selection and skip to next parameter Note At any time you can press the ESC button to exit the Setup mode and resume normal operat...

Страница 13: ...E CHECK button to display the year month day hour minutes and seconds Checking Sampling Rate During normal operation hold the Sampling check button to view the selected sampling rate Data Logging Data...

Страница 14: ...seconds to resume normal operation Note The meter will scan for a SD memory card when a first manual recording is attempted If a card is not inserted or the card is defective the meter will flash CArD...

Страница 15: ...low for an Excel sample file Place Date Time Value Unit 1 19 07 10 0 05 02 42 5 dB 2 19 07 10 0 05 04 45 3 dB 3 19 07 10 0 05 06 52 dB 4 19 07 10 0 05 08 58 5 dB 5 19 07 10 0 05 10 58 1 dB 6 19 07 10...

Страница 16: ...s microphone adapter 7 Power on the sound calibrator R8090 If the measured value is not within 94 0 2 dB adjust the meter s calibration potentiometer until the display matches the output of the calibr...

Страница 17: ...kickstand 2 Remove the battery cover 3 Replace 6 x 1 5V AA batteries 4 Secure the battery cover with the two 2 Phillips screws Applications Preventative Maintenance Industrial Machinery Equipment Con...

Страница 18: ...ake sure to exit that function by holding down the REC button before attempting to turn your meter off How long can I record for Battery life will depend on a number of factors Quality of Battery Whet...

Страница 19: ...be defective because of improper material or workmanship under normal use and maintenance REED Instruments total liability is limited to repair or replacement of the product REED Instruments shall no...

Страница 20: ...ts com Please visit www REEDINSTRUMENTS com for the most up to date manuals datasheets product guides and software Product specifications subject to change without notice All rights reserved Any unaut...

Страница 21: ...Test Measure with Confidence Check out our latest Products REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Страница 22: ...REED Instruments 1 877 849 2127 info reedinstruments com www reedinstruments com...

Отзывы: