14
3.
Multi-point measurement
: This method involves performing a
number of measurements within a circle having a maximum diameter
of about 1.18" (30mm) (see image below). Take the minimum
indicated value as the thickness of the material.
Menu Options
The menu function controls the settings and functions of the meter. To
enter the menu, press the
button to highlight the menu option on the
LCD and press the
button.
System Setup
1. While in the menu, highlight the "System Setup" option and press the
button to enter this menu.
2.
Press the
and
buttons to scroll through the "System Setup" menu.
3. When the required setting as shown below, is highlighted, press the
button to select it and toggle between units.
Note
: To set brightness, press
button to enter the setting, and use
the
and
buttons to adjust the brightness.
4. Once completed, press the
button to save and exit.
System Setup Menu Settings are as follows:
1.
Measurement Units
: Metric and Imperial
2.
Receiving Gain
: LOW (resolution of 0.1mm) and HIGH (resolution
of 0.01mm). LOW is mainly used for measuring coarse material with
high scatter and small sound absorption, such as cast aluminum,
cast copper, and other metallic parts.
3.
Minimum Capture Measurement
: OFF and ON
4.
2-Point Calibration
: OFF and ON
5.
Auto Down
: Power-saving mode ON (default)
continued...
www.
.com
1.800.561.8187