10
Changing Units of Measure
1. Press and hold the UNIT button, while in RMS measurement mode.
2. The LCD display will toggle between the following units:
ACC
m/s
2
ACC g
VEL
mm/s
VEL
cm/s
ACC
ft/s
2
VEL
inch/s
3. Release the UNIT button to select the displayed unit.
4. The selected unit will be saved as the new default.
Zero Adjustment
The zero adjustment is intended to remove small offsets caused by
environmental changes.
1. Ensure the vibration sensor is properly secured to the meter.
2. Set the measurement function to "Acceleration" (m/s², g, ft/s²).
3. Keep the vibration sensor motionless so no signal is picked up.
4. Press and hold the
LOG
button for approx. 5 seconds or until a zero
value is displayed.
Note:
The zero adjustment function only works when the measured value
is less than 10 digits.
Max Hold Reset
While in the Max Hold function, press and hold the
LOG
button for
3 seconds to clear the current max hold value.