8
7.
Place the presser foot onto the test piece while keeping the foot
parallel to the surface.
8.
Apply enough force to ensure proper contact between the presser
foot and the test piece.
9.
Keep in place and Hold for approx. 2 seconds.
10. The LCD will display the measured maximum value and will hold
this value.
11. Press the
ZERO
button to take the next test measurement.
12. Follow steps 6 through 11 until the selected number of test
measurements have been reached.
13. When completed, the meter will emit 2 beeps and automatically
display the average measurement value of the test session.
Auto Power Off
To preserve battery life, the meter is programmed to turn itself off after
10 minutes of inactivity.
Zero Calibration
Press the
ZERO
button to perform a zero calibration when required.
Single Point Calibration
1.
Place the indenter onto a flat test piece of glass.
2.
Apply enough force to make firm contact between the glass and the
presser foot.
3.
The readings on the display should be between 99.7 and 101.
4.
If the reading is not within the above range, hold the
button until
"CAL" appears on the display while continuing to keep pressure
against the glass.
5.
When the reading has stabilized, press the
button to display "0.0"
which confirms that calibration is complete.
6.
Repeat steps 1 through 3 to confirm that the reading is now
within range.
7.
Repeat single point calibration if reading is still out of range.