WARNING: Disconnect all power to the unit before 

installing Plug-in cards.

Adding Option Cards

The PAX and MPAX series meters can be fitted with up to three optional plug-

in cards. The details for each plug-in card can be reviewed in the specification 

section below. Only one card from each function type can be installed at one time. 

The function types include Setpoint Alarms (PAXCDS), Communications 

(PAXCDC), and Analog Output (PAXCDL). The plug-in cards can be installed 

initially or at a later date.

COMMUNICATION CARDS (PAXCDC)

A variety of communication protocols are available for the PAX and MPAX 

series. Only one of these cards can be installed at a time. 

Note: For Modbus 

communications use RS485 Communications Output Card and configure 

communication (



) parameter for Modbus.

PAXCDC10 - RS485 Serial (Terminal)

PAXCDC30 - DeviceNet

PAXCDC1C - RS485 Serial (Connector)

PAXCDC50 - Profibus-DP

PAXCDC20 - RS232 Serial (Terminal)

PAXUSB00 - USB (Mini B)

PAXCDC2C - RS232 Serial (Connector)

SERIAL COMMUNICATIONS CARD

Type

: RS485 or RS232

Communication Type

: RLC Protocol (ASCII), Modbus RTU, and Modbus 

ASCII

Isolation To Sensor & User Input Commons

: 500 Vrms for 1 min.

Working Voltage: 50 V.   Not Isolated from all other commons. 

Baud

: 300 to 38,400

Data

: 7/8 bits

Parity

: No, Odd or Even

Bus Address

: Selectable 0 to 99 (RLC Protocol), or 1 to 247 (Modbus 

Protocol), Max. 32 meters per line (RS485)

Transmit Delay

: Selectable for 0 to 0.250 sec (+2 msec min)

DEVICENET™ CARD

Compatibility

: Group 2 Server Only, not UCMM capable

Baud Rates

: 125 Kbaud, 250 Kbaud, and 500 Kbaud

Bus Interface

: Phillips 82C250 or equivalent with MIS wiring protection per 

DeviceNet™ Volume I Section 10.2.2.

Node Isolation

: Bus powered, isolated node

Host Isolation

: 500 Vrms for 1 minute (50 V working) between DeviceNet™ 

and meter input common.

PROFIBUS-DP CARD

Fieldbus Type:

 Profibus-DP as per EN 50170, implemented with Siemens 

SPC3 ASIC

Conformance:

 PNO Certified Profibus-DP Slave Device

Baud Rates: 

Automatic baud rate detection in the range 9.6 Kbaud to 12 Mbaud

Station Address

: 0 to 125, set by rotary switches.

Connection: 

9-pin Female D-Sub connector

Network Isolation: 

500 Vrms for 1 minute (50 V working) between Profibus 

network and sensor and user input commons. Not isolated from all other 

commons.

PAXUSB PROGRAMMING CARD

Type

: USB Virtual Comms Port

Connection

:

 

Type mini B

Isolation To Sensor & User Input Commons

: 500 Vrms for 1 min.

Working Voltage: 50 V. Not Isolated from all other commons.

Baud Rate

: 300 to 19.2k

Unit Address

: 0 to 99; only 1 meter can be configured at a time

SETPOINT CARDS (PAXCDS)

The PAX and MPAX series has 4 available setpoint alarm output plug-in 

cards. Only one of these cards can be installed at a time. (Logic state of the 

outputs can be reversed in the programming.) These plug-in cards include:

PAXCDS10 - Dual Relay, FORM-C, Normally open & closed

PAXCDS20 - Quad Relay, FORM-A, Normally open only

PAXCDS30 - Isolated quad sinking NPN open collector

PAXCDS40 - Isolated quad sourcing PNP open collector

DUAL RELAY CARD

Type

: Two FORM-C relays

Isolation To Sensor & User Input Commons

: 2000 Vrms for 1 min.

Working Voltage: 240 Vrms

Contact Rating

One Relay Energized: 5 amps @ 120/240 VAC or 28 VDC (resistive load), 

1/8 HP @120 VAC, inductive load

Total current with both relays energized not to exceed 5 amps

Life Expectancy

: 100 K cycles min. at full load rating. External RC snubber 

extends relay life for operation with inductive loads

QUAD RELAY CARD

Type

: Four FORM-A relays

Isolation To Sensor & User Input Commons

: 2300 Vrms for 1 min.

Working Voltage: 250 Vrms

Contact Rating

One Relay Energized: 3 amps @ 240 VAC or 30 VDC (resistive load), 1/10 

HP @120 VAC, inductive load

Total current with all four relays energized not to exceed 4 amps

Life Expectancy

: 100 K cycles min. at full load rating. External RC snubber 

extends relay life for operation with inductive loads

QUAD SINKING OPEN COLLECTOR CARD

Type

: Four isolated sinking NPN transistors. 

Isolation To Sensor & User Input Commons

: 500 Vrms for 1 min.

Working Voltage: 50 V.    Not Isolated from all other commons. 

Rating

: 100 mA max @ V

SAT

 = 0.7 V max. V

MAX

 = 30 V

QUAD SOURCING OPEN COLLECTOR CARD

Type

: Four isolated sourcing PNP transistors.

Isolation To Sensor & User Input Commons

: 500 Vrms for 1 min.

Working Voltage: 50 V.    Not Isolated from all other commons. 

Rating

: Internal supply: 24 VDC ± 10%, 30 mA max. total

External supply: 30 VDC max., 100 mA max. each output

ALL FOUR SETPOINT CARDS

Response Time

: See update rates step response specification; add 6 msec 

(typical)for relay card

O

pTiOnal

 p

lug

-

in

 O

uTpuT

 C

ards

4

a

CCessOries

UNITS LABEL KIT (PAXLBK)

Each meter has a units indicator with backlighting that can be customized 

using the Units Label Kit. The backlight is controlled in the programming.

PROGRAMMING SOFTWARE

The Crimson

®

 2 (SFCRM2) software is a Windows

®

 based program for 

configuring and updating the firmware of the PAXDP meter from a PC. Using 

the software makes programming the PAXDP meter easier and allows the user 

to save the PAXDP database in a PC file for future use. The software is available 

as a free download from Red Lion’s website.

The first time Crimson 2 software is run from the File menu, select “New” to 

display a dialog and select the PAXDP. The screen will display icons that 

represent the various programming sections of the PAXDP. Double-click on an 

icon to configure the programming parameters pertaining to the selection. Tool 

Tip help is available for each of the  program parameters. A PAX serial plug-in 

card or PAX USB programming card is required to program the meter using the 

software.

When communicating with Crimson 2 software, the PAXDP must be set in 

default configuration type of: 

Communications Type: MODBUS RTU

Baud Rate: 38400

Data Bit: 8

ParityBit: no

Meter Unit Address: 247 

Содержание PAXDP

Страница 1: ...eadout of totalized flow calculate service intervals of motors or pumps etc The totalizer can also accumulate batch operations The meter has four setpoint outputs implemented on Plug in option cards T...

Страница 2: ...S30 Quad Setpoint Sourcing Open Collector Output Card PAXCDS40 PAXCDC RS485 Serial Communications Card with Terminal Block PAXCDC10 Extended RS485 Serial Communications Card with Dual RJ11 Connector P...

Страница 3: ...Response Time 20 msec max Logic State Jumper selectable for sink source logic 12 TOTALIZER Function Time Base second minute hour or day Batch Can accumulate gate input display from a user input Time...

Страница 4: ...lude PAXCDS10 Dual Relay FORM C Normally open closed PAXCDS20 Quad Relay FORM A Normally open only PAXCDS30 Isolated quad sinking NPN open collector PAXCDS40 Isolated quad sourcing PNP open collector...

Страница 5: ...rature and provides good air circulation Placing the unit near devices that generate excessive heat should be avoided The bezel should be cleaned only with a soft cloth and neutral soap product Do NOT...

Страница 6: ...to the unit as possible Loop the cable through the core several times or use multiple cores on each cable for additional protection Install line filters on the power input cable to the unit to suppres...

Страница 7: ...In order to preserve the safety of the meter application the sensor input common must be suitably isolated from hazardous live earth referenced voltages or input common must be at protective earth gro...

Страница 8: ...device between appropriate User Input terminal and User Comm DSP 8 8 8 8 8 A C B PAR F1 F2 RST A 1 S P S P S 2 3 P P4 S Display Readout Legends Optional Custom Units Overlay Setpoint Alarm Annunciato...

Страница 9: ...listed on the Parameter Value Chart following the programming section ALTERNATING SELECTION DISPLAY In the module description sections which follow the dual display with arrows appears for each progra...

Страница 10: ...less than the band value the filter engages again This allows for a stable readout but permits the display to settle rapidly after a large process change The value of the band is in display units ind...

Страница 11: ...mA 0 with 20 mA the desired high Display Value The calculations stop at the limits of the Input Range Jumper position 6 For input levels beyond the last programmed Input Value the meter extends the D...

Страница 12: ...B Display will normally show the Relative unless switched by this function The Absolute is a gross value based on Module 1 DSP and INP entries without the Display Offset Value The Absolute display is...

Страница 13: ...ECTIONS Select main or alternate setpoints Reset Setpoint 1 Alarm 1 Reset Setpoint 2 Alarm 2 Reset Setpoint 3 Alarm 3 Reset Setpoint 4 Alarm 4 Reset Setpoint 3 4 Alarm 3 4 Reset Setpoint 2 3 4 Alarm 2...

Страница 14: ...s to be viewed and modified This Programming Mode can be locked with a security code and or user input When locked and the PAR key is pressed the meter enters a Quick Programming Mode In this mode the...

Страница 15: ...present MAX value for the entered delay time the meter will capture that display value as the new MAX reading A delay time helps to avoid false captures of sudden short spikes to sec MAX CAPTURE ASSIG...

Страница 16: ...ut permits the display to settle rapidly after a large process change The value of the band is in display units independent of the Display Decimal Point position A band setting of 0 keeps the digital...

Страница 17: ...Totalizer Decimal Point parameter TOTALIZER HIGH ORDER DISPLAY When the total exceeds 5 digits the front panel annunciator flashes if assigned to A B or C display In this case the meter continues to t...

Страница 18: ...with unbalanced hysteresis Deviation high with unbalanced hysteresis Deviation low with unbalanced hysteresis Inside band with unbalanced hysteresis Outside band with unbalanced hysteresis Lower Total...

Страница 19: ...0 0 allows the meter to update the alarm status per the response time listed in the Specifications When the output logic is this becomes on time delay Any time accumulated at power off resets during p...

Страница 20: ...not be used The PAXCDC1 RS485 or PAXCDC2 RS232 card should be used instead COMMUNICATIONS TYPE RLC Protocol ASCII Modbus RTU Modbus ASCII The Communication Type factory settings must be changed from...

Страница 21: ...udes read only registers then only the write registers will change 4 If the write value exceeds the register limit see Register Table then that register value changes to its high or low limit FC17 Rep...

Страница 22: ...o word N A N A N A Read Only 40003 Input B Relative Value Hi word N A N A N A Read Only 40004 40005 40006 40007 40009 40011 40013 40017 40015 40019 40008 40010 40012 40014 40018 40016 40020 Calculatio...

Страница 23: ...ess specifier the next character is the command character 3 The next character is the Register ID This identifies the register that the command affects The P command does not require a Register ID cha...

Страница 24: ...ot change the corresponding output mode U abcde e Analog Output d SP4 c SP3 b SP2 a SP1 Example VU00011 places SP4 and Analog in manual Analog Output Register AOR ID W This register stores the present...

Страница 25: ...RMAT Data is transferred from the meter through a serial communication channel In serial communications the voltage is switched between a high and low level at a predetermined rate baud rate using ASC...

Страница 26: ...E ANALOG UPDATE TIME to to to Enter the source for the analog output to retransmit Relative net Input Value The Relative Input Value is the Absolute Input Value that includes the Display Offset Value...

Страница 27: ...ve selections to be calibrated At each prompt use the PAX arrow keys to adjust the external meter display to match the selection being calibrated When the external reading matches or if this range is...

Страница 28: ...K Power level power connections Module 3 programming NO DISPLAY REMEDIES PROBLEM For further assistance contact technical support at the appropriate company numbers listed PARAMETER VALUE CHART Progra...

Страница 29: ...RE ASSIGNMENT CALCULATION CONSTANT VALUE CALCULATION FUNCTION MIN CAPTURE DELAY TIME INPUT B OFFSET VALUE MAX CAPTURE DELAY TIME INPUT A OFFSET VALUE CALCULATION FILTER BAND DISPLAY INTENSITY LEVEL US...

Страница 30: ...SP t LO t LO AS HI AS HI t SCFAC ASIN dECPt dECPt tbASE Locut P uP PAr INP A CALC INP b tYPE dAtA bAUd SPNt HILO tot Addr dELAY YES Abru OPt NO RLC MODE ONLY ASn n Scaling Style FUNCTION KEYS USER INP...

Страница 31: ...31 This page intentionally left blank...

Страница 32: ...Buyer its employees or sub contractors are or may be to any extent liable including without limitation penalties imposed by the Consumer Product Safety Act P L 92 573 and liability imposed upon any p...

Отзывы: