red lion PAXDP Скачать руководство пользователя страница 27

27

27

d-LEV

Display

Intensity Level

PAR

9-FCS

Pro

Factory 

Service Code

COdE

6.9 mOdule 9 - f

aCTOry

 s

erviCe

 O

peraTiOns

 (



)

PARAMETER MENU

Enter the desired Display Intensity Level (0-15) by 

using the arrow keys. The display will actively dim or 

brighten as the levels are changed. This parameter also 

appears in Quick Programming Mode when enabled.

DISPLAY INTENSITY LEVEL



The meter has been fully calibrated at the factory. 

Scaling to convert the input signal to a desired display 

value is performed in Module 1. If the meter appears to be 

indicating incorrectly or inaccurately, refer to 

Troubleshooting before attempting to calibrate the meter.

When recalibration is required (generally every 2 years), it should only be 

performed by qualified technicians using appropriate equipment. Calibration 

does not change any user programmed parameters. However, it may affect the 

accuracy of the input signal values previously stored using the Apply (



Scaling Style.

Calibration may be aborted by disconnecting power to the meter before 

exiting Module 9. In this case, the existing calibration settings remain in effect.

CALIBRATION









RESTORE FACTORY DEFAULTS

Use the arrow keys to display 

 

 and press 

PAR

The meter will display 



 and then return to 

 

Press 

DSP

 key to return to Display Mode. This will 

overwrite all user settings with the factory settings.

ANALOG OUTPUT CARD CALIBRATION

Before starting, verify that the precision voltmeter (voltage output) or current 

meter (current output) is connected and ready. Perform the following procedure:

1. Use the arrow keys to display 

 

 and press 

PAR

.

2. Use the arrow keys to choose 



 and press 

PAR

.

3. Using the chart below, step through the five selections to be calibrated. At 

each prompt, use the PAX arrow keys to adjust the external meter display to 

match the selection being calibrated. When the external reading matches, or 

if this range is not being calibrated, press 

PAR

.

4. When 



 appears remove the external meters and press 

PAR

 twice.

SELECTION

EXTERNAL METER

ACTION



0.00

Adjust if necessary, press 

PAR



4.00

Adjust if necessary, press 

PAR



20.00

Adjust if necessary, press 

PAR



0.00

Adjust if necessary, press 

PAR



10.00

Adjust if necessary, press 

PAR

WARNING: Calibration of this meter requires a signal source with an 

accuracy of 0.01% or better and an external meter with an accuracy of 

0.005% or better.

Before starting, verify that the Input Ranger Jumper is set for the range to be 

calibrated. Also verify that the precision signal source is connected and ready. 

Allow a 30 minute warm-up period before calibrating the meter. 



 and 

PAR 

can be chosen to exit the calibration mode without any changes taking place.

Then perform the following procedure:

1. Use the arrow keys to display (

 

) and press 

PAR

.

2. Choose the input channel/range to be calibrated by using the arrow keys and 

press 

PAR

. (



 and 

PAR

 can be chosen to exit the calibration mode without 

any changes taking place.)

3. When the zero range limit appears on the display, apply the appropriate:

- Voltage range: dead short applied

- Current range: open circuit

4. Press 

PAR

 and the top range limit will appear on the display after 

approximately 1 second.

5. With the top range limit on the display, apply the appropriate:

- Voltage range: 10 VDC

- Current range: 20 mADC

6. Press 

PAR

 and 





 will appear on the display after approximately 1 

second.

7. When 



 appears, press 

PAR

 twice.

8. If the meter is not field scaled, then the input display should match the value 

of the input signal.

9. Repeat the above procedure for each input range to be calibrated.

INPUT CALIBRATION

Содержание PAXDP

Страница 1: ...eadout of totalized flow calculate service intervals of motors or pumps etc The totalizer can also accumulate batch operations The meter has four setpoint outputs implemented on Plug in option cards T...

Страница 2: ...S30 Quad Setpoint Sourcing Open Collector Output Card PAXCDS40 PAXCDC RS485 Serial Communications Card with Terminal Block PAXCDC10 Extended RS485 Serial Communications Card with Dual RJ11 Connector P...

Страница 3: ...Response Time 20 msec max Logic State Jumper selectable for sink source logic 12 TOTALIZER Function Time Base second minute hour or day Batch Can accumulate gate input display from a user input Time...

Страница 4: ...lude PAXCDS10 Dual Relay FORM C Normally open closed PAXCDS20 Quad Relay FORM A Normally open only PAXCDS30 Isolated quad sinking NPN open collector PAXCDS40 Isolated quad sourcing PNP open collector...

Страница 5: ...rature and provides good air circulation Placing the unit near devices that generate excessive heat should be avoided The bezel should be cleaned only with a soft cloth and neutral soap product Do NOT...

Страница 6: ...to the unit as possible Loop the cable through the core several times or use multiple cores on each cable for additional protection Install line filters on the power input cable to the unit to suppres...

Страница 7: ...In order to preserve the safety of the meter application the sensor input common must be suitably isolated from hazardous live earth referenced voltages or input common must be at protective earth gro...

Страница 8: ...device between appropriate User Input terminal and User Comm DSP 8 8 8 8 8 A C B PAR F1 F2 RST A 1 S P S P S 2 3 P P4 S Display Readout Legends Optional Custom Units Overlay Setpoint Alarm Annunciato...

Страница 9: ...listed on the Parameter Value Chart following the programming section ALTERNATING SELECTION DISPLAY In the module description sections which follow the dual display with arrows appears for each progra...

Страница 10: ...less than the band value the filter engages again This allows for a stable readout but permits the display to settle rapidly after a large process change The value of the band is in display units ind...

Страница 11: ...mA 0 with 20 mA the desired high Display Value The calculations stop at the limits of the Input Range Jumper position 6 For input levels beyond the last programmed Input Value the meter extends the D...

Страница 12: ...B Display will normally show the Relative unless switched by this function The Absolute is a gross value based on Module 1 DSP and INP entries without the Display Offset Value The Absolute display is...

Страница 13: ...ECTIONS Select main or alternate setpoints Reset Setpoint 1 Alarm 1 Reset Setpoint 2 Alarm 2 Reset Setpoint 3 Alarm 3 Reset Setpoint 4 Alarm 4 Reset Setpoint 3 4 Alarm 3 4 Reset Setpoint 2 3 4 Alarm 2...

Страница 14: ...s to be viewed and modified This Programming Mode can be locked with a security code and or user input When locked and the PAR key is pressed the meter enters a Quick Programming Mode In this mode the...

Страница 15: ...present MAX value for the entered delay time the meter will capture that display value as the new MAX reading A delay time helps to avoid false captures of sudden short spikes to sec MAX CAPTURE ASSIG...

Страница 16: ...ut permits the display to settle rapidly after a large process change The value of the band is in display units independent of the Display Decimal Point position A band setting of 0 keeps the digital...

Страница 17: ...Totalizer Decimal Point parameter TOTALIZER HIGH ORDER DISPLAY When the total exceeds 5 digits the front panel annunciator flashes if assigned to A B or C display In this case the meter continues to t...

Страница 18: ...with unbalanced hysteresis Deviation high with unbalanced hysteresis Deviation low with unbalanced hysteresis Inside band with unbalanced hysteresis Outside band with unbalanced hysteresis Lower Total...

Страница 19: ...0 0 allows the meter to update the alarm status per the response time listed in the Specifications When the output logic is this becomes on time delay Any time accumulated at power off resets during p...

Страница 20: ...not be used The PAXCDC1 RS485 or PAXCDC2 RS232 card should be used instead COMMUNICATIONS TYPE RLC Protocol ASCII Modbus RTU Modbus ASCII The Communication Type factory settings must be changed from...

Страница 21: ...udes read only registers then only the write registers will change 4 If the write value exceeds the register limit see Register Table then that register value changes to its high or low limit FC17 Rep...

Страница 22: ...o word N A N A N A Read Only 40003 Input B Relative Value Hi word N A N A N A Read Only 40004 40005 40006 40007 40009 40011 40013 40017 40015 40019 40008 40010 40012 40014 40018 40016 40020 Calculatio...

Страница 23: ...ess specifier the next character is the command character 3 The next character is the Register ID This identifies the register that the command affects The P command does not require a Register ID cha...

Страница 24: ...ot change the corresponding output mode U abcde e Analog Output d SP4 c SP3 b SP2 a SP1 Example VU00011 places SP4 and Analog in manual Analog Output Register AOR ID W This register stores the present...

Страница 25: ...RMAT Data is transferred from the meter through a serial communication channel In serial communications the voltage is switched between a high and low level at a predetermined rate baud rate using ASC...

Страница 26: ...E ANALOG UPDATE TIME to to to Enter the source for the analog output to retransmit Relative net Input Value The Relative Input Value is the Absolute Input Value that includes the Display Offset Value...

Страница 27: ...ve selections to be calibrated At each prompt use the PAX arrow keys to adjust the external meter display to match the selection being calibrated When the external reading matches or if this range is...

Страница 28: ...K Power level power connections Module 3 programming NO DISPLAY REMEDIES PROBLEM For further assistance contact technical support at the appropriate company numbers listed PARAMETER VALUE CHART Progra...

Страница 29: ...RE ASSIGNMENT CALCULATION CONSTANT VALUE CALCULATION FUNCTION MIN CAPTURE DELAY TIME INPUT B OFFSET VALUE MAX CAPTURE DELAY TIME INPUT A OFFSET VALUE CALCULATION FILTER BAND DISPLAY INTENSITY LEVEL US...

Страница 30: ...SP t LO t LO AS HI AS HI t SCFAC ASIN dECPt dECPt tbASE Locut P uP PAr INP A CALC INP b tYPE dAtA bAUd SPNt HILO tot Addr dELAY YES Abru OPt NO RLC MODE ONLY ASn n Scaling Style FUNCTION KEYS USER INP...

Страница 31: ...31 This page intentionally left blank...

Страница 32: ...Buyer its employees or sub contractors are or may be to any extent liable including without limitation penalties imposed by the Consumer Product Safety Act P L 92 573 and liability imposed upon any p...

Отзывы: