VOLTAGE CALIBRATION

1. Connect a precision Voltage Source with an output range of -10.000 to 60.000

mV and an accuracy of 0.03% or better to the TC- (#5) and TC+ (#3)
terminals. Cold junction calibration must be performed AFTER voltage
calibration.

2. Enter the Program Menu and step through the Program Menu list by pressing

the “SEL” button until 

CAL

is displayed.

3. Press the “blank” button to change 

no

to 

YES

. Press “SEL”.

4. The display shows 

00000. 

Use the “blank” button to increment the access

code to 

00006.

Press and hold “SEL”.

5. After entering the correct access code, the unit will display 

IN1

. Follow the

table below for calibration sequence.The unit displays 

CALC

for about 3

seconds after pressing the “SEL” button before proceeding to the next input
point.

6. After voltage calibration is complete, the unit will return to the 

CAL

program

item. Exit programming and verify various points from -10.00 to 60.00
mVDC using the Voltage Calibration Check procedure to ensure that the
calibration is correct.

7. Repeat calibration if necessary.

COLD JUNCTION CALIBRATION

This procedure must be performed AFTER an accurate voltage calibration.

The ambient temperature at which this calibration should be performed is 15 to
35 °C (59 to 95 °F). A Temperature indicator with 0.1 °C resolution and
accuracy is required to perform this calibration.
1. Enter Programming Menu, and verify the following:

tyPE

= probe being connected to the unit.

CJC

YES

dISP

0.0

°

C

OFSEt

0

Make any necessary changes to programming.

2. Connect a thermocouple probe of known accuracy and type (J, K, E, T, or N)

to the CUB4TC. The probe should match the one selected in programming.

3. Connect a reference temperature probe to measuring end of the CUB4TC

thermocouple.The two probes should be shielded from air movement and
allowed sufficient time (15 minutes minimum) to equalize in temperature.
(As an alternative, the CUB4TC thermocouple probe may be placed in a
calibration bath of known temperature.)

4. Compare the unit display with reference temperature indicator (or calibration

bath). If the displayed unit temperature does not equal the reference probe
temperature, calculate the CJ error as follows:
CJ Error = reference probe temperature - CUB4TC displayed temperature 

5. Enter the Program Menu and step through the Program Menu list by pressing

the “SEL” button until 

CAL

is displayed.

6. Press the “blank” button to change 

no

to 

YES

. Press “SEL”.

7. The display shows 

00000. 

Use the “blank” button to increment the access

code to 

00003.

Press and hold “SEL”.

8. After entering the correct access code, the display will alternate between 

CJC

and the current cold junction compensation value.

9. If the current CJC value is 0.0 °C or 59.9 °C, set the 

CJC

value to the current

ambient temperature, exit calibration, and repeat the CJC calibration
procedure.

10. Otherwise, the new value to be entered is the sum of the current value and

the CJ Error recorded previously.

11. Exit the Programming Menu and repeat the calibration from Step 4 to verify

accuracy.

Voltage Calibration Wiring

DISPLAY

PARAMETER

ACTION

IN1

-10.000 mV

Apply -10.000 mV, wait 5 sec., press 

SEL

IN2

0.000 mV

Apply  0.000 mV, wait 5 sec., press 

SEL

IN3

10.000 mV

Apply 10.000 mV, wait 5 sec., press 

SEL

IN4

20.000 mV

Apply 20.000 mV, wait 5 sec., press 

SEL

IN5

30.000 mV

Apply 30.000 mV, wait 5 sec., press 

SEL

IN6

40.000 mV

Apply 40.000 mV, wait 5 sec., press 

SEL

IN7

50.000 mV

Apply 50.000 mV, wait 5 sec., press 

SEL

IN8

60.000 mV

Apply 60.000 mV, wait 5 sec., press 

SEL

Cold Junction Calibration Wiring

Содержание CUB4RT

Страница 1: ...TC Backlight version 0 15 C Volt CJC error if powered from other than 12 VDC May be compensated with offset Example 24 VDC supply 24V 12V 0 15 C V 1 8 C Enter 1 8 C user offset 3 TEMPERATURE EFFECTS C...

Страница 2: ...o 1400 F 1 9 WHITE YELLOW R 0 to 1768 C 32 to 3214 F 4 5 BLACK WHITE S 0 to 1768 C 32 to 3214 F 4 5 BLACK WHITE B 200 to 1820 C 300 to 3308 F 9 1 540 C 4 5 540 C GREY NONE E 200 to 787 C 328 to 1448 F...

Страница 3: ...ges FLPrG and or FLCAL will be periodically flashed on the display FLPrG indicates that non calibration data is corrupt This flashing display will remain active even if power is removed and reapplied...

Страница 4: ...ber or press the SEL button to wrap around to the LSD After the desired value is on the display press and hold the SEL button for three seconds to enter the value and return to the main menu CAL YES N...

Страница 5: ...lay will read 00 000 to 20 000 ohms All other RTD types with the jumper in position B will cause the display to read 000 00 to 400 00 ohms Apply 105 000 mV wait 5 sec press SEL 105 000 mV IN8 Apply 90...

Страница 6: ...reference temperature probe to measuring end of the CUB4TC thermocouple The two probes should be shielded from air movement and allowed sufficient time 15 minutes minimum to equalize in temperature A...

Страница 7: ...the hole in the mounting clip 4 Install the unit through the panel cut out until the front bezel flange contacts the panel 5 Slide the mounting clip over the rear of the unit until the mounting clip i...

Страница 8: ...after No warranties expressed or implied are created with respect to The Company s products except those expressly contained herein The Customer acknowledges the disclaimers and limitations contained...

Отзывы: