red lion CUB4RT Скачать руководство пользователя страница 1

1

!

PROGRAMMABLE INPUTS

CUB4TC - Thermocouple types T, E, J, K, R, S, B, N, or mV
CUB4RT - RTD types Pt385, Pt392, Ni672, Cu427

!

PROGRAMMABLE TEMPERATURE OFFSET

!

SELECTABLE F or C WITH 1 or 0.1 RESOLUTION

!

F OR C DISPLAY ANNUNCIATORS

!

CONFORMS TO ITS-90 TEMPERATURE SCALE 

!

COLD JUNCTION COMPENSATION (Enable/Disable) CUB4TC

!

0.48 INCH (12.2 mm) HIGH DIGITS

!

5 DIGIT LCD, POSITIVE REFLECTIVE OR NEGATIVE

TRANSMISSIVE WITH RED BACKLIGHT

!

NEMA 4X/IP65 SEALED FRONT BEZEL

DESCRIPTION

These CUB4 temperature meters are 9 to 26 VDC powered, microprocessor

based indicators, with a selectable decimal point. Models are available for use
with RTD or Thermocouple sensors. The CUB4TC offers cold junction
compensation and a mVDC display mode with a range of -10.00 to 60.00 mV.
Programmable features include temperature type, Fahrenheit or Celsius display,
input filtering range, and user offset. 

SAFETY SUMMARY

All safety related regulations, local codes, and instructions that appear in the

manual or on equipment must be observed to ensure personal safety and to
prevent damage to either the instrument or equipment connected to it. If
equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.

SPECIFICATIONS

1. 

DISPLAY:

Reflective LCD or Transmissive LCD with red backlight, 5 digit

display 0.48"(12.2 mm) height, F or C annunciator.

2. 

POWER:

9 to 26 VDC @ 65 mA with backlight, 25 mA without backlight.

Reverse polarity protected.
Must use the MLPS or a Class 2 or SELV rated power supply.

CUB4TC Backlight version

: 0.15 

°

C/Volt CJC error if powered from other

than 12 VDC. May be compensated with offset.
Example: 24 VDC supply, (24V-12V)*0.15 °C/V = 1.8 °C, 

Enter -1.8 °C user offset.

3.

TEMPERATURE EFFECTS:

CUB4RT Span Drift

: 100 ppm/°C Max

CUB4TC Span Drift

: 150 ppm/°C Max.

Zero Drift

: < 1 

µ

V/°C Max.

4. 

ENVIRONMENTAL CONDITIONS:

Operating Temperature Range

: 0 to 50 °C

Storage Temperature Range

: -30 to 85 °C

Operating and Storage Humidity

: 85% max. relative humidity

(non-condensing) from 0 to 50 °C.

Altitude

: Up to 2000 meters (6561 feet)

5.

MAIN SENSOR INPUT RANGES AND ACCURACIES

Isolation

: The sensor input and excitation, if applicable are not isolated from

the power supply.

Response Time

: 500 msec

Failed Sensor Display

: OPEN

Overrange/Underrange Input

: OLOL/ULUL

Overrange/Underrange Display

: “…..”/”-…..”

Normal Mode Rejection

: 40 dB @ 50/60 Hz

Common Mode Rejection

: 100 dB @ DC to 60 Hz

Accuracies reflect ALL errors @ 12 VDC supply and 15 minute warm-up,
except probe errors and lead resistance effects.

MODEL CUB4RT/TC - INTELLIGENT TEMPERATURE INDICATORS

CAUTION: 

Read complete instructions prior to installation and

operation of the unit.

DIMENSIONS  In inches (mm)

Note: Recommended minimum clearance (behind the panel) for

mounting clip installation is 2.1" (53.3) H x 3.0" (76.2) W.

PANEL CUT-OUT

Bulletin No.  CUB4RT/TC-X
Drawing No.  LP0583
Effective  10/04

Tel +1 (717) 767-6511
Fax +1 (717) 764-0839
www.redlion.net

UL Recognized Component,
File # E179259

DESCRIPTION

PART NUMBER

MLPS

CUB4RT

CUB4TC

RTD Indicator, Reflective LCD

Thermocouple Indicator, Reflective LCD

CUB4RT00

CUB4TC00

Micro Line/Sensor Power Supply

RTD Indicator, Red Backlit Transmissive LCD

Thermocouple Indicator, Red Backlit Transmissive LCD

MLPS1000

CUB4RT20

CUB4TC20

ORDERING INFORMATION

MODEL NO.

Содержание CUB4RT

Страница 1: ...TC Backlight version 0 15 C Volt CJC error if powered from other than 12 VDC May be compensated with offset Example 24 VDC supply 24V 12V 0 15 C V 1 8 C Enter 1 8 C user offset 3 TEMPERATURE EFFECTS C...

Страница 2: ...o 1400 F 1 9 WHITE YELLOW R 0 to 1768 C 32 to 3214 F 4 5 BLACK WHITE S 0 to 1768 C 32 to 3214 F 4 5 BLACK WHITE B 200 to 1820 C 300 to 3308 F 9 1 540 C 4 5 540 C GREY NONE E 200 to 787 C 328 to 1448 F...

Страница 3: ...ges FLPrG and or FLCAL will be periodically flashed on the display FLPrG indicates that non calibration data is corrupt This flashing display will remain active even if power is removed and reapplied...

Страница 4: ...ber or press the SEL button to wrap around to the LSD After the desired value is on the display press and hold the SEL button for three seconds to enter the value and return to the main menu CAL YES N...

Страница 5: ...lay will read 00 000 to 20 000 ohms All other RTD types with the jumper in position B will cause the display to read 000 00 to 400 00 ohms Apply 105 000 mV wait 5 sec press SEL 105 000 mV IN8 Apply 90...

Страница 6: ...reference temperature probe to measuring end of the CUB4TC thermocouple The two probes should be shielded from air movement and allowed sufficient time 15 minutes minimum to equalize in temperature A...

Страница 7: ...the hole in the mounting clip 4 Install the unit through the panel cut out until the front bezel flange contacts the panel 5 Slide the mounting clip over the rear of the unit until the mounting clip i...

Страница 8: ...after No warranties expressed or implied are created with respect to The Company s products except those expressly contained herein The Customer acknowledges the disclaimers and limitations contained...

Отзывы: