TROUBLESHOOTING
Please use the generated error code to detect and follow the
instructions to fault resolution within the following overview of
error descriptions. If the error occurs repeatedly, please contact the
technical support at the RECK Company or an authorized local partner.
Overview of most important error messages
The most important error codes are listed below. Using the displayed
error codes from the display of the device, you will find information
and solutions for the problem in this chart.
If the displayed error code is not listed and a restart or a short
disconnection from the mains does not solve the problem, please
contact the technical support department at the RECK Company or an
authorized MOTOmed representative.
page 78
page 78
Error code
A000140
»overheating«
A000150
»Motor
blocking«
A000202
»Security
stop«
MOTOmed reaction
Error message
appears during the
training.
Error message
appears during the
training.
Error appears after
pushing the start /
stop button.
Hints for troubleshooting
Disconnect the mains cable and
allow for a
device cool down of
at least 15 minutes. Thereafter
reconnect the device.
Check if the pedals are moving
freely. Restart the device by
pushing the red start/stop button
again.
In case of high muscle tune, you
can try increasing the motor
power.
Restart the device by pushing
the red start/stop button again.
Alternatively, you can disconnect
and reconnect by pulling the
mains cable.
Possible cause
Overheating of
device due to:
-
Blocking of pedal
crank.
-
Continuous
training at a high
muscle tone.
-
Defect
temperature
sensor.
-
With activated
SpasmControl
feature more
than 15 spasms
are detected
subsequently and
consecutively.
-
If the motor
is blocking for
longer than
60 seconds
with turned off
SpasmControl.
The stop button
was pushed for 3
seconds or more.