Step
Description
Image
3
Remove the absorbent pad contaminated with buffer oil
.
4
Replace the O-ring on the oil purge container (see
for details on handling O-rings)
5
Replace the absorbent pad
Roll one side of a clean pad and insert it into the container's
central compartment, letting the rest of the pad out through
the slit on the side
Wrap the remaining pad around the container going
clockwise once
If any length of the pad is left, loosen it so that there is only
one full layer wrapped around the container (any additional
layers will make it difficult to insert)
6
Reinstall the container
Hold the pad so that it does not unravel
Insert the container back into the instrument
Turn the red handle clockwise one full rotation
7
Return the battery end-cap back onto the instrument (see
Opening and closing the instrument
Dispose the contaminated pad in accordance with local
regulations.