Maintenance
Marathon
Series
FA/FR
47
The
relay
is
controlled
by
the
temperature
selected
on
the
display.
If
any
failsafe
code
appears
on
the
display,
the
relay
changes
to
the
“abnormal”
state.
The
exception
is
the
“dirty
window”
condition.
This
causes
the
relay
to
change
state,
leaving
a
normal
numerical
temperature
output.
The
dirty
window
is
detected
in
either
1
‐
color
or
2
‐
color
mode.
Error Code
0 – 20 mA Output
4 – 20 mA Output
no error
according to temperature according to temperature
ECHH
21 to 24 mA
21 to 24 mA
ECUU
0 mA
2 to 3 mA
EIHH
21 to 24 mA
21 to 24 mA
EIUU
0 mA
2 to 3 mA
EUUU
0 mA
2 to 3 mA
EHHH
21 to 24 mA
21 to 24 mA
EAAA
0 mA
2 to 3 mA
Table
11:
Current
Output
Values
in
accordance
to
an
Error
If
two
errors
occur
simultaneously,
the
higher
priority
error
is
the
one
that
is
presented
on
the
LED’s
digital
and
analog
outputs.
For
example,
in
2
‐
color
mode,
if
the
internal
ambient
is
too
high
and
the
attenuation
is
too
high,
the
unit
outputs
EIHH
on
the
LED’s
and
digital
output
and
21
mA
on
the
analog
output.
However,
since
2
‐
color
wide
band
and
narrow
band
temperatures
may
all
be
presented
simultaneously
through
RS485,
their
over
and
under
range
conditions
are
independent.
Examples
of
failsafe
conditions:
1.
One
‐
color
temperature
is
selected
for
display
on
the
LED’s.
Two
‐
color
temperature
is
transmitted
in
burst
mode.
Wide
band
temperature
is
under
range.
Two
‐
color
temperature
is
999°C.
Outputs:
Display:
EUUU
RS485:
C
T0999
Analog:
2
to
3
mA
Relay:
abnormal
state
2.
Two
‐
color
temperature
is
selected
for
display
on
LED’s.
All
three
temperatures
are
transmitted
in
burst
mode.
Two
‐
color
temperature
is
1021°C.
Wide
band
temperature
is
703°C.
Narrow
band
temperature
is
685°C.
Attenuation
is
above
95%,
the
“dirty
window”
threshold.
Outputs:
Display:
1021
RS485:
C
T1021
W0703
N0685
Analog:
scaled
to
temperature,
between
4
and
20
mA
Relay:
abnormal
state