4.4.8 Important tips and tricks for the test with the RAH Module 10                           Page 47 

 

3. Directly Altering the harmonisation time regarding a stored resonance 
value 

 

After storing a resonance value, it is always possible to give it an individual 
harmonisation time directly. 

 
Therefore you press the key: 

 

 

 
 

Afterwards, the RAH-programmes can also be related with an individual time 
(fig. 1). In front of the double point, the hours are entered, the minutes 

behind it. Later on, when you store all found RAH-programmes on a RAH 
‘Green Card’, the individual time will be stored under this programme. 

 

 

1

 

4. Very important: Deleting the resonance memory 

 
 Always, when a new patient is going to be analysed, the resonance memory 

has to be deleted. After switching on the device, the memory is empty, and 
while testing with RAH-programmes and their harmonisation times, the 
memory gets filled. 

 
During the RAH testing, the resonance memory can be deleted by pressing  

 
the key              

             

 

for

 

a longer time.

 

The device confirms this  

 
with a long beep-signal.  

 
When an RAH ‘Green Card’ is read in for testing, the total resonance memory 

will be deleted automatically beforehand, as the user would get confused 
otherwise. The same is valid for the test with test protocols. 
 

 
 

 
 

 

Содержание Rayocomp PS 10

Страница 1: ...information advice and suggestions for solving problems this book cannot replace a visit to an alternative practitioner or a doctor orientated to natural healing Also we must point out that the classi...

Страница 2: ...he Mode Key 14 4 1 5 Operating the Integrated Polariser and Sensor 18 4 1 6 Basic Information for RAH Green Card 20 4 1 7 Error Display 21 4 2 Menu Item Direct 22 4 2 1 Testing and Harmonising a Funda...

Страница 3: ...lues 52 4 5 6 Activating and Deactivating Modules 53 4 5 7 Storing Fundamental Frequency Values on Memory Cards 54 4 5 8 Displaying Memory Card Information 55 4 5 9 Providing Read Protection for a Mem...

Страница 4: ...2 Conformity Declaration Page 4...

Страница 5: ...nd accessories are secured inside the case In case you intend to send back the PS10 please always include the appurtenant charging set 3 2 Technical Data Weight of the Rayocomp PS 10 1 6 kg Weight inc...

Страница 6: ...the charging cable plugged in When using the detectors especially the pointed detector for acupuncture points be sure to handle them with care risk of eye injury 3 4 Guarantee The warranty period runs...

Страница 7: ...accompanied by short descriptions in order to simplify the assignment 3 6 1 Front View 1 2 4 3 6 7 5 Front view of the Rayocomp PS 10 Caption 1 Display 2 Keyboard 3 Three exit hubs for connecting the...

Страница 8: ...ear view of the Rayocomp PS 10 Caption 1 ON OFF switch 2 Charging socket 3 Charging plug which is inserted in the charging socket for the charging process 4 Interface for program updates 5 Sticker ind...

Страница 9: ...ams for example 7 Stop key for interrupting an automatic run or for deleting the function 8 N key of the polariser to switch the polariser to neutral 9 key of the polariser to switch on the bipolar fu...

Страница 10: ...vice is due 4 In the menu item F1 you can read the remaining total harmonising time here 5 This displays the polarisation type which is currently set 6 Here you can read which menu item has been selec...

Страница 11: ...area as bio resonance as developed by Paul Schmidt Meanwhile there is sufficient literature on the market from which you can take the application possibilities of the method For example we recommend...

Страница 12: ...to the rear of the device see chapter 3 6 2 4 The charging unit indicates the various charging processes When the LED turns green the charging process is complete and you can use the device again 5 Pl...

Страница 13: ...ge Er 29 the power pack must be recharged before you can work with the device Then the device switches over to the menu item Direct and is immediately ready for action The display shows It is not nece...

Страница 14: ...in the device The RAH programme which is also found under the Menu Item F1 is shown only if it has effectively been activated The RAH module is marked by the indication RAH in the display After a sho...

Страница 15: ...is not pressed again for a short moment the PS10 will automatically start the input mask allowing you to enter the BnPS programmes By pressing the Mode key again the Rayocomp PS 10 will switch over t...

Страница 16: ...tem Test RAH will be shown on the display after the menu item Test BnPS By pressing on the key mode once again this menu item can be selected If the key mode isn t pressed once more after having selec...

Страница 17: ...ocomp PS 10 changes to the menu item Setup and shows on the display The Rayocomp PS 10 changes over to the menu item Info and displays If you press the key no you will move back to the menu item Direc...

Страница 18: ...the keyboard is rather bothering when performing tests we recommend using the button of the sensor to carry out the functions of the integrated polariser right from the start Basically the use of the...

Страница 19: ...ind a summary of the sensor functions green Wooden handle Changing the polarisation Next fundamental frequency value Previous fundamental frequency value in the internal memory of the PS 10 Storing a...

Страница 20: ...and the PS 10 add automatically well proven harmonising times Furthermore the values stored on the RAH Green Card can be read in for a further test This saves a lot of time and avoids entering errors...

Страница 21: ...e an entry or storage 2 Two short sounds The key you are pressing has no current function 3 Two long sounds This signals a severe error The device will supply an error code In the appendix chapter 5 8...

Страница 22: ...n the menu item Direct to the Rayocomp PS 10 The polariser integrated in the device is set to N Test with the sensor to find out whether you achieve a rotation left or right motion or a linear movemen...

Страница 23: ...e 56 00 After switching on your device you arrive in the menu item Direct The display shows Please note that entering commas is in this case optional However the comma can be helpful when entering one...

Страница 24: ...ry card slot 3 Wait until the data have been loaded then unplug the card 4 The automatic harmonisation sequence begins At first the fundamental frequency values stored on the RAH Green Card are set up...

Страница 25: ...d to the development of the control plan The set up within the harmonising programs always follows the same pattern The first fundamental frequency values are the pre control values taken from the con...

Страница 26: ...tomatically set up the fundamental frequency values of the pre control first followed by the values for suppressing scar interference Before giving detailed examples for the entry let us summarise imp...

Страница 27: ...rmonise Activate the key for until the menu item F1 appears Now you can enter the first number of the harmonising program 1 51 Press the following keys The display shows Note Here you can see the numb...

Страница 28: ...want to harmonise each fundamental frequency value for one minute and thirty seconds you only have to enter 01 30 For your convenience the Rayocomp PS 10 calculates the total harmonising time at once...

Страница 29: ...p PS 10 and press the following key The Rayocomp PS 10 can now read the values stored on the memory card and harmonise with them Tip 1 You can combine the values of an integrated program with those of...

Страница 30: ...m numbers from the special brochures Actuate the mode key several times until you see the menu item F1 BnPS in the display Note The Analysis and Harmonising system RAH will only work when there is no...

Страница 31: ...to determine the treatment course with these programs please press the following keys Now the display will look like this These values indicate the length of the harmonising time of the currently sele...

Страница 32: ...cted so far Confirm the entry of the harmonisation time with the OK button The following fig shows the new display Now you can enter the second program The 2 in front of the harmonising time indicates...

Страница 33: ...play shows the progress After 100 is achieved the polariser switches over to and the timer for harmonising starts counting backward The countdown of the total harmonising time proceeds in steps of one...

Страница 34: ...rt value The previous program is only called up again if you repeat this step Testing a RAH Harmonising Program with the Rayotensor The harmonising programs of the Rayocomp PS 10 cannot be applied for...

Страница 35: ...already integrated in the storage of the device and no longer need to be entered After the first frequency value has been tested you must switch to the following program value Do this by means of the...

Страница 36: ...re special compilations of fundamental frequency values You can use them if module 6 test programs is activated in your Rayocomp PS 10 Important information The fundamental frequency values tested wit...

Страница 37: ...he fundamental frequency values in disorder at the moment then to store and harmonise them With the bio resonance devices according to Paul Schmidt it is possible to carry out a resonance test in a ve...

Страница 38: ...esting fundamental frequency values to find resonances 4 4 2 Testing with Memory Card Values With the Rayocomp PS 10 it is very easy to save fundamental frequency values on a memory card The values st...

Страница 39: ...ted programs the device also contains a compilation of test programs which are very frequently required They are listed below once more for your convenience Display Program name P 04 Electro magnetic...

Страница 40: ...efficiently in the analysis of energetic disfunctions Using the example of electro smog tests we concretise it as follows If there is a general force of electro smog RAH No 04 00 you can narrow it on...

Страница 41: ...and detail levels under the current selected programme On the highest RAH programme level 45 programmes are available fig 6 The programme 04 00 is the fourth fig 5 of 45 programmes of the highest lev...

Страница 42: ...is the first fig 5 of 3 programmes of the second level For the test the polariser of the Rayocomp PS 10 is automatically switched on N fig 3 There are currently no programmes selected that have to be...

Страница 43: ...me 04 20 which can be found itself under the main group 04 00 The whole structure of the RAH programmes can be analysed with the Rayocomp PS 10 as described in this chapter and in case there is any ne...

Страница 44: ...th the disease pattern will be recommended to you directly for the testing at the push of one button Sometimes this can be more than 100 different RAH programmes An RAH programme with an available tes...

Страница 45: ...st them Please insert a programmed RAH Green Card into the Rayocomp PS 10 and press the following key The RAH programmes stored on the card will now be read into the Rayocomp PS 10 and are then availa...

Страница 46: ...programme is loaded completely and is ready for testing When an RAH programme was stored as a resonance a long beep signal sounds after successful storage The acoustic signals are very helpful because...

Страница 47: ...Green Card the individual time will be stored under this programme 1 4 Very important Deleting the resonance memory Always when a new patient is going to be analysed the resonance memory has to be de...

Страница 48: ...their harmonisation times indicated on the display fig 1 will be stored 1 Storing the RAH programme numbers can also be performed under the menu item F1 This programme item is best suited for entering...

Страница 49: ...up number to address the desired setup function For exmple if you want to enter the setup number 10 please press 1 then 0 and confirm your entry with the OK key Further actions are described in the fo...

Страница 50: ...off at night if you are harmonising while you are resting Your sleep is not disturbed by the bright light of the illumination in the background On delivery the standard setup of the Rayocomp PS 10 is...

Страница 51: ...rent error This setup function enables or disables the signal Enter the setup number 12 and you can add one of the following numbers 1 The acoustic signal is switched on 0 The acoustic signal is switc...

Страница 52: ...an be stored but not for the single frequencies If an automatic programme sequence is started by the means of an RAH Green Card the default for the single frequencies can be regulated by the pre setti...

Страница 53: ...ou can select the number of the module intended for activation or deactivation Please confirm each input with the OK key Here is a list of the module numbers Module No Designation 01 Setup precision o...

Страница 54: ...The frequency compilations can be stored on a normal memory card or a Green Card Insert the memory card in your device and enter the setup number 30 Every entered fundamental frequency number will no...

Страница 55: ...shown in the display with the 200 at the same time e g 34 200 Insert the memory card and enter the setup number 40 The display may show 3 2 1 Caption 1 This number indicates the amount of values alre...

Страница 56: ...e enter 1 and confirm your entry with the OK key Please keep in mind the read protection of a memory card can only be reversed by deleting it see chapter 4 5 10 4 5 10 Providing Write Protection for a...

Страница 57: ...stage and confirm with the OK key 4 5 12 Initialising a Memory Card Setup Number 44 If you should come upon a memory card with illegible sticker you can test the card with this setup function and rein...

Страница 58: ...n perfect working order with this function After you have entered the setup number 50 and confirmed your entry with the OK key the Rayocomp PS 10 starts a trial run You can read the progress of the tr...

Страница 59: ...art No Designation 01 2010 Setup precision of 0 05 02 2020 54 integrated harmonising programs 03 2030 14 Acupuncture oscillation therapy 04 2040 22 Programmes according to Dr Ulrich 05 2050 9 special...

Страница 60: ...em 7 This symbolises the load condition of the battery pack 8 This number stands for the construction month of the device in this case December 9 This is the construction year of the device It was the...

Страница 61: ...n of the IF keys is exactly the same Enter the desired fundamental frequency value in the menu item Direct Confirm the entry with OK Then press the desired IF key until you hear a signal Set the desir...

Страница 62: ...Testing with range values P 02 Test with memory card values P 03 Test with integrated programs P 04 Electro magnetic pollution test P 05 Geopathy test P 06 Check up test P 07 Bacteria and viruses tes...

Страница 63: ...1 19 Lung P 1 20 Hormone glands P 1 21 Spinal cord P 1 22 Larynx P 1 23 Diaphragm P 1 24 Pleura P 1 25 Peritoneum P 1 26 Male sexual organs P 1 27 Female sexual organs P 1 28 Ureter P 1 29 Stomach P 1...

Страница 64: ...Colon channel P 2 03 Stomach channel P 2 04 Spleen channel P 2 05 Heart channel P 2 06 Channel of the small Intestines P 2 07 Bladder channel P 2 08 Kidney channel P 2 09 Liver channel P 2 10 Channel...

Страница 65: ...ummary of the Integrated Special Programs Display Program name P 3 01 Check up P 3 02 Scar interference suppression P 3 03 Bacteria viruses P 3 04 Parasites P 3 05 Global screen system P 3 06 Water ve...

Страница 66: ...s program P 4 08 Female hormone program P 4 09 Male hormone program P 4 10 After care program P 4 11 P 4 12 Tinnitus program P 4 13 Constants of physics program P 4 14 Clearing program P 4 15 Spiritua...

Страница 67: ...egrated harmonising programs 03 2030 14 Acupuncture oscillation therapy 04 2040 22 Programs according to Dr Ulrich 05 2050 9 special programs 06 2070 11 Test programs 08 2080 Rayonex Analysis and Harm...

Страница 68: ...26 Maximum capacity of the memory full 27 This module has already been enabled temporarily 28 No Green Card inserted 29 Very low current Charge battery pack 30 Service is due 31 This program group is...

Страница 69: ...d successively 87 No values on the memory card 88 No valid memory card inserted Please check or if necessary reinitialise the card see Setup Item 40 or 44 respectively 89 More than 500 values selected...

Страница 70: ...6 Private Documents Page 70 6 Private Documents...

Страница 71: ...6 Private Documents Page 71 6 Private Documents...

Страница 72: ...Rayonex Schwingungstechnik GmbH Sauerland Pyramiden 1 57368 Lennestadt Tel 49 2721 6006 0 Fax 49 2721 6006 66 Internet www rayonex de E mail info rayonex de FB 146 Rev28 Stand 15 September 2011...

Отзывы: