Wuhan Raycus Fiber Laser Technologies Co., Ltd
User Guide of 1500W CW laser
Set the water temperature of the cooling system correctly
according to the ambient temperature. If the water
temperature is set too high, the laser will not work properly.
If the water temperature is set too low, condensation will
occur inside the laser or the laser output cable, which will
cause irrecoverable damage to the laser.
Before staring the laser, the cooling system must work
properly and the water temperature reaches a suitable
temperature.
3.4 Installation rules
1) Place the product in an appropriate position, immobilize it if necessary.
2) Check if the power supply has the correct voltage (laser model and
power voltage refers to product parameters in Table 5 ), and the earth
line is connected, and ensure the earth line is firmly connected.
3) Connect the power cable and control cable to the product when power
supply is OFF.
4) Connect the cooling system to the laser and output fiber cables according
to the water inlet and outlet signs.
5) Please check the laser output head and do the necessary cleaning before
installing it in the device. If you find that there is dust or foreign objects
that cannot be cleaned, please contact Raycus in time, and do not
continue installation and operation for the time being.
6) Prevent the delivery cable from treading, pinching or excessive bending
during installation.
17