SENSITIVITY KNOB ADJUSTMENTS 

The Sensitivity Knob is used to compensate for varying types of sludge density. 

 

 

 Knob 

Setting 

 Probe 

Attitude 

   Sensitivity 

 

 

 

        1   

 

Ignoring the smallest solids 

     

      Low 

 

 

 

        5   

 

Detecting all solids 

 

 

      High 

 
Note: Sensitivity range set at factory.  Range modification instructions available upon request. 

IN THE FIELD 

SLUDGE INTERFACE DETECTION

1.  Rotate Sensitivity Knob to position [1] low sensitivity. 
2. Lower 

the 

SID

 probe all the way to the bottom of the tank.  NOTE: If needed, increase sensitivity just 

enough to light up all the lightbars.  THIS IS THE SENSITIVITY BASELINE FOR THIS TANK 

3.  Slowly raise the 

SID

 probe out of the sludge blanket until most or all of the lightbars go out.  At this level, 

you have detected the sludge blanket interface. 

4. Observe 

the 

SID

 probe cable depth marker to determine the depth of blanket. 

IN THE FIELD

FINE TUNING 

IN THE FIELD

• 

Now that you’ve established the baseline sensitivity for this tank, you can fine-tune the 

SID

 by increasing 

the sensitivity above the baseline. 

 

• 

At the baseline sensitivity setting determined in the previous step, the probe is exhibiting an “all or nothing”
reading.  At sensitivities higher than your baseline, you can detect the ragged edge on top of the sludge 
blanket.   

 

• 

By lowering and raising the 

SID

 probe in and out of the sludge blanket interface at various sensitivity 

settings above your baseline, you will get a feel for the 

SID’s

 “personality” and how it reacts to the 

particular influents at your plant. 

MAINTENANCE

IN THE SHOP

• 

Replace all 6 batteries as a group with new AA alkaline batteries of the same type and brand 

• 

Switch 

SID

 power switch-off when not in use to conserve batteries 

• 

Store 

SID

 for long periods with the batteries removed 

• 

Store 

SID

 in cool dry place and out of direct sunlight 

• 

Clean probe sensor eyes with mild detergent 

Replacement Parts 

S-10203-20   20-foot Cable Assembly with Sludge Sensor Probe 
S-10203-30   30-foot Cable Assembly with Sludge Sensor Probe 
 

1 year limited warranty.   
 
Other Raven Environmental Products process control instruments: 

 

F-10300 

Bench top Centrifuge 

(includes six 19mm Centrifuge Tubes)

 

B-10101-19 

19 mm Centrifuge Tubes 

(conical bottom with percentage markings) 

B-10101-17 

17 mm Centrifuge Tubes 

(conical bottom with percentage markings)

 

C-10102 Settleometer 

Kit

 (Measures sludge settling characteristics) 

 