background image

17 

 

The potentiometers 
with which we 
calibrate the output of 
the load cells, are 
mechanical parts 
therefore, higher risk 
components  

Move the board and but pressure with fingers on the 
potentiometers while looking at the display to see if it 
reacts. Do not touch the contact itself. 

Not linear 

Check if it is load cells 
or indicator 

Load cells or indicator are very rarely the cause of this 
problem. Easiest way to check is by changing the 
indicator temporarily. If problem is not solved when 
changing the indicator, the problem is the load cell, 
cable or mechanics  

Check cable 

Very rarely the cause. Maybe in a lift truck.  

Instability  With no load it 

is most of the 
time humidity, 
bad connection 
or component r 
bad shield. 

Check for humidity 

Check for water marks on the indicator board or load 
cell connections (potentiometers). 

Check the indicator. 

Sometimes the indicator will show a weight when load 
cells are disconnected. If you do this and the indicator 
becomes more stable, it is most likely elsewhere in 
the system. 
Check visually for traces of oxidation. If found heating 
the solder contacts can solve the problem. 

Check cables. In 
warehouse and lift 
truck the cable is 
working all the time 
when following the 
lifting movement. It 
may be worn or 
damaged. Changing 
temperatures and 
chemicals have an 
effect on the lifetime of 
a cable. 

Bad connections will cause changes when moving the 
scale.  

Bend and move the cable briskly especially where the 
cable is moving continuously when lifting. While doing 
so, look at the display to see if it reacts to the 
movements.  

The potentiometers 
with which we 
calibrate the output of 
the load cells are 
mechanical parts and 
are sensitive to 
humidity, shocks and 
vibration. 

Move the board and but pressure with fingers on the 
potentiometers while looking at the display to see if it 
reacts. Do not touch the contact itself. 
  

Check the load cells.   If connected independently to the indicator, it can be 

checked which one is unstable and which one is not. 

With load  

Check mechanics. 

 

Function 
error 

No reaction 
when pushing 
keys 

Check the touch panel  Test can be done by making short cut on connection 

of the touch panel to simulate a key being pressed. 
Check for wear of broken contacts in the flat cable 
going to the indicator board 

Lock up 

Take out the batteries and replace to see if it starts up 
afterwards. 

Содержание 100 Series

Страница 1: ...t the fact that this RAVAS product is 100 recyclable on the basis that the parts are processed and disposed off in the right manner More information can be found on our website www ravas com Rev 20201102 Printing typographical errors and model changes reserved ...

Страница 2: ...on Manual 7 3 1 Accurate Weighing 3 2 Touch Panel Display 3 3 Net Tare Gross 3 4 On Off 3 5 Error Messages 3 6 Zero and Tare Functions 3 7 Manual Tare 3 8 Summing 3 9 Total and reset 4 Trouble Shooting 16 5 Calibration 19 6 Parameter Settings 21 Contact Us salesoffice ravas com www ravas com ...

Страница 3: ...being completed 9 It is the sole responsibility of the purchaser to train their own employees in the proper use and maintenance of this equipment 10 Do not operate this unit unless you have been fully trained of its capabilities 11 Do not use the weighing system in potentially explosive areas 12 Do not carry passengers with the truck 13 Do not weld or make changes to the weighing system without co...

Страница 4: ...osition Thread the chain through the hole in the triangle and through the hole in the axle Place the handle bar onto the triangle and insert the bolts Tighten the bolts firmly Push the silver part A on the outside of the pump downwards At the same time insert the chain into the open side of the slot B on the inside of the pump A B ...

Страница 5: ... the system takes place through 4 AA batteries When used normally the batteries will last for about 1 year When the voltage level of the batteries is running low the display will show LO BA When the batteries are completely empty the weighing system shuts off To replace the batteries ...

Страница 6: ...ANUAL INDEX PAGE 3 1 Accurate Weighing 7 3 2 Touch Panel Display 8 3 3 Net Tare Gross 8 3 4 On Off 9 3 5 Error Messages 10 3 6 Zero and Tare Functions 12 3 7 Manual Preset Tare 13 3 8 Summing 14 3 9 Total and reset 15 ...

Страница 7: ...7 3 1 Accurate Weighing 2 ...

Страница 8: ...8 MIN 3 2 Touch Panel Indicator 3 3 Net Tare Gross 1 2 3 ...

Страница 9: ...9 3 4 On Off ...

Страница 10: ...10 3 5 Error Messages max ...

Страница 11: ...11 ...

Страница 12: ...12 3 6 Zero and Tare Functions ...

Страница 13: ...13 3 7 Manual Tare Entry ...

Страница 14: ...14 3 8 Summing ...

Страница 15: ...15 3 9 Total and Reset ...

Страница 16: ...e if they can interfere with the load cells With the forks lifted half way up the brackets for the loading wheels may touch the fork shoe By taking off the fork shoe Scratches will show if it does and where it does Check if bolts are loose Check the load cells If one is broken or gives more or less signal than the others the scale will give different reading depending how it is loaded To be sure t...

Страница 17: ...ating the solder contacts can solve the problem Check cables In warehouse and lift truck the cable is working all the time when following the lifting movement It may be worn or damaged Changing temperatures and chemicals have an effect on the lifetime of a cable Bad connections will cause changes when moving the scale Bend and move the cable briskly especially where the cable is moving continuousl...

Страница 18: ...P 2 Scale is overloaded Take load from scale If there is no load do the same checks as you do with HELP 3 and 7 Help 3 or 7 Load cell signal too high or too low Check cables for damage Move the cable while looking at display to see if indicator reacts Measure load cells to see if they are fine Check the excitation signal of the indicator Help 4 Out of zero range Zero calibration needed ...

Страница 19: ...ALIBRATION Deletion resetting to zero of earlier calibrated points Push the PT key for about 18 seconds this may be 8 seconds on older versions The display will go blank until calibration mode has been reached The indicator shows the value of the first calibration point the lowest indication bar on the left of the display is flashing Use the and keys to see the three earlier programmed values on t...

Страница 20: ...g the value per segment by pressing the 0 T key until the proper weight has been entered When each segment has been corrected the lower indication bar will flash again Confirm the entered weight by pressing the 0 T key for 3 seconds The display counts down from AF 08 to AF 00 the first calibration point has now been set Press the key The higher indication bar will start flashing and the indicator ...

Страница 21: ...he 0 T key The indicator shows the current value of parameter P 01 Use keys and to change the current value Confirm with 0 T The indicator shows P 02 Parameter P 02 can now be changed in the same way as P 01 To scroll through the parameters press the key until the desired parameter is reached To jump back to P 01 quickly press and release the on off key while a parameter number is shown in the dis...

Страница 22: ...22 ...

Страница 23: ...23 ...

Отзывы: