CALIBRATION GUIDE
rapidcount X 300 P
4.4 Calibration of CISIR transmitted light
4.4.1
Once you have followed the steps
1. to 4.
and you arrive in the
„SENSOR ADJUST“
interface, select
„CISIR TRN ADJUST“
op-
tion by pressing the
„CLR“
and
„ADD“
key
to navigate the menu. To select the
„CISIR
TRN ADJUST“
option, press the
„RESET/
ENTER“
key.
4.4.2
When the following messages are display-
ed, it means the machine is ready to adjust
the
CISIR transmitted light
.
4.4.3
Open the back cover fi rst, then insert the
pink
calibration paper, making sure the
CISIR transmitted light
are fully covered.
Close the back cover.
4.4.4
Press the
„RESET/ENTER“
key, the machi-
ne will automatically adjust the sensors.
Then the following screen will be shown.
The
EMIT
-Value should be within
105~180
,
and the
REC
-Value should be within
53~73
.
After adjustment, remove the
pink
calibra-
tion paper. You may turn off the device.
Pinkes
Kalibrierungspapier!