CALIBRATION GUIDE 

rapidcount X 300 P

4.4 Calibration of CISIR transmitted light

4.4.1 

Once you have followed the steps 

1. to 4. 

and you arrive in the 

„SENSOR ADJUST“ 

interface, select 

„CISIR TRN ADJUST“

 op-

tion by pressing the 

„CLR“

 and 

„ADD“

 key 

to navigate the menu. To select the 

„CISIR 

TRN ADJUST“

 option, press the 

„RESET/

ENTER“

 key.

4.4.2

When the following messages are display-
ed, it means the machine is ready to adjust 
the 

CISIR transmitted light

.

4.4.3

Open the back cover fi rst, then insert the 

pink

 calibration paper, making sure the 

CISIR transmitted light

 are fully covered. 

Close the back cover.

4.4.4

Press the 

„RESET/ENTER“

 key, the machi-

ne will automatically adjust the sensors.
Then the following screen will be shown. 
The 

EMIT

-Value should be within 

105~180

and the 

REC

-Value should be within 

53~73

After adjustment, remove the 

pink

 calibra-

tion paper. You may turn off  the device.