4.1.1

Once you have followed the steps 

1. to 4.

 

and you arrive in the 

„SENSOR ADJUST“

 

interface, select 

„COUNTER ADJUST“

 

option by pressing the 

„CLR“

 and 

„ADD“

 

key to navigate the menu. To select the 

„COUNTER ADJUST“

 option, press the 

„RESET/ENTER“

 key.

4.1.2

When the following messages are display-
ed, it means the machine is ready to adjust 
the counting 

IR sensors

4.1.3

To adjust the sensors, open the back cover 
fi rst, then insert the 

pink

 calibration pa-

per, making sure the counting 

IR sensors 

are fully covered. Close the back cover.

4.1.4

Press the 

„RESET/ENTER“

 key, the machi-

ne will automatically adjust the sensors. 
Then the following screen will be shown. 
The 

EMIT

-Value should be within 

108~200

, and the 

REC

-Value should be 

within 

44~50

. After the adjustment, re-

move the 

pink

 calibration paper. Press 

the 

„CUR“

 key to return to the 

„SENSOR 

ADJUST“ 

interface. 

[EN]

CALIBRATION GUIDE 

rapidcount X 300 P

Pink

calibration paper!