4.1.1
Once you have followed the steps
1. to 4.
and you arrive in the
„SENSOR ADJUST“
interface, select
„COUNTER ADJUST“
option by pressing the
„CLR“
and
„ADD“
key to navigate the menu. To select the
„COUNTER ADJUST“
option, press the
„RESET/ENTER“
key.
4.1.2
When the following messages are display-
ed, it means the machine is ready to adjust
the counting
IR sensors
.
4.1.3
To adjust the sensors, open the back cover
fi rst, then insert the
pink
calibration pa-
per, making sure the counting
IR sensors
are fully covered. Close the back cover.
4.1.4
Press the
„RESET/ENTER“
key, the machi-
ne will automatically adjust the sensors.
Then the following screen will be shown.
The
EMIT
-Value should be within
108~200
, and the
REC
-Value should be
within
44~50
. After the adjustment, re-
move the
pink
calibration paper. Press
the
„CUR“
key to return to the
„SENSOR
ADJUST“
interface.
[EN]
CALIBRATION GUIDE
rapidcount X 300 P
Pink
calibration paper!