![Rational TC 61/101 E Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/rational/tc-61-101-e/tc-61-101-e_operator-and-installation-manual_3852982014.webp)
V06
- 14 -
Electrical connection
• Follow the installation instructions and the infor-
mation on the rating plate when connecting the
unit
• The units must be connected to a standard supply
line in conformity with your local regulations.
• Comply with all local regulations and standards!
• We recommend the use of an earth leakage
circuit breaker (30mA).
• On-site installation: provide accessible all-pole
disconnection device with a minimum of a 3 mm
contact gap.
• A permanent electrical connection must be
provided for the units.
The thermo cabinet (TC) will be delivered with a
power cord without plug. The connection box is
located on the bottom of the TC.
pic 3 A
Connect the power cord in the following way:
PE (yellow/green wire)
earth
N (blue wire)
neutral
L1 (brown wire)
phase
Also there is a connection bolt for the earth con-
nection, located at the rear of the left front leg.
Connect the TC with a suitable cable to the earth
bonding system
pic. 3 B
GB
3
A
B
Electrical connection