background image

Ratek WBx000D Series Water Bath - User Manual 

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

 

 

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ 

Page 10 of 21

 

If the water level falls below this minimum level during operation, the water bath should be powered off, disconnected from the 
mains electrical supply and the water level topped up with cold water. Be aware that this will result in the rapid decrease in the 
water bath water temperature and the effect of this in relation to the current procedure being carried out must be considered. 
 

 

Using The Water Bath 

 
 

Filling The Water Tank 

 

 

 

CAUTION

 

It is important that the mesh platform always remain in place during operation to provide protection from the heat generated 
by the heating element. Operating the water bath without the mesh platform creates a hazardous situation sufficient to cause 
burns or scalding if contact with the element is made. 

 

IMPORTANT: Do not fill or operate the water bath unless the mesh platform has been fitted. 
 

 

 

 

-

 

For models fitted with a drain tap (WB1100D, WB1700D), ensure the tap is in the closed position before filling. 

 

-

 

The water bath should be filled by slowly pouring cold water (below 25° Celsius) into the end of the water bath furthest 
from the control panel until the desired working level is reached. 

 

-

 

Avoid splashing as much as possible.

 

 

 

 

Connecting Power 

 

 

 

CAUTION

 

The equipment must only be used with a protective earth power socket. The earth contact provides protection to the user and 
the equipment. If you do not have a protective earth power socket, or you are unsure as to whether you have a protective 
earth power socket, 

do not

 connect the equipment. In such cases you should consult your workplace administrator or electrical 

maintenance staff to determine if a protective earth power socket is available. 
 
The mains power supply must be rated to match the power requirement as identified on the product identification label on the  
rear panel. A surge protected power outlet is strongly recommended as it provides some protection for the equipment in areas 
of poor electrical quality as well as providing some protection against lightning strikes. The equipment should be operated on a 
good, reliable supply of power at all times. 

 

Note: The water bath should not be operated on the same electrical circuit as other high voltage household appliances such as 
fridges, clothes dryers, washing machines or other continuous operation high voltage devices. These types of devices can 
create power fluctuations that are undesirable for electrically sensitive equipment. Consult your workplace administrator or 
electrical maintenance staff if you are unsure. 

 

IMPORTANT : Use of an incorrect power supply will void the product warranty waive all liability for any and all damage caused by 
such use. If you are unsure about the rating of your power supply, consult your workplace administrator or electrical 
maintenance staff to determine if your power supply is suitable for use with this product before connecting the power lead. 

 

 

 

CAUTION 

Regularly check the mains power lead condition over the life of the product, and do not operate the equipment if you suspect 
there is damage to any part of the equipment or the mains power lead. 
 
Do not operate the equipment if you suspect the power lead has been stretched, over-extended or damaged in any way.

  

 

 

 

-

 

Ensure your hands are dry before connecting power. 

-

 

Insert the female IEC end of the cable firmly into the socket on the rear of the waterbath and ensure a good fit. 

-

 

Insert the plug end of the mains power supply lead firmly into a properly rated, protective earthed wall mounted power 
supply outlet. 

 
If there are double-adapters or oversized DC power packs causing obstruction of the mains power lead plug, these should first 
be removed. 

 

 

Содержание WB 000D Series

Страница 1: ...rsion 1 0 29 02 2016 English www ratek com au 60 Wadhurst Dve Boronia Victoria 3155 Australia P 613 9887 2161 F 613 9887 2163 E sales ratek com au All Rights Reserved Made in Australia User Manual pri...

Страница 2: ...uct to ensure it has not been damaged in transit Any damage should be reported immediately to the responsible carrier If the product is damaged in any way re pack the product into the supplied packagi...

Страница 3: ...eral Operation 4 OPERATOR RESPONSIBILITY SAFETY CONSIDERATIONS 4 Conditions of Operation 4 Safety Labels And Markings 5 OPERATING PROCEDURES 6 Identification of Controls Functions 6 Safety Warnings 7...

Страница 4: ...e water bath is operated via a front panel interface consisting of heating temperature control buttons plus digital display and an on off switch located at the rear of the bath These controls allow th...

Страница 5: ...s or questions relating to operator or user safety please contact the appropriate Ratek Service department before installing and operating the unit Contact details are provided in this User Manual Saf...

Страница 6: ...ing a control or component of the water bath will be used to identify it The figure below indicates all key controls and components of the water bath with their corresponding numerical element labeled...

Страница 7: ...place safety regulations in effect CAUTION If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired CAUTION The equipment must on...

Страница 8: ...emptying NEVER operate the equipment if you believe it is damaged in any way NEVER operate the water bath if the mains power supply cable is damaged in any way NEVER use any sharp or metal objects ne...

Страница 9: ...into the water bath ALWAYS use plastic stainless steel or plastic powder coated tubes racks and accessories in the water bath This will assist to prevent corrosion and thus help to extend the useful l...

Страница 10: ...ated to match the power requirement as identified on the product identification label on the rear panel A surge protected power outlet is strongly recommended as it provides some protection for the eq...

Страница 11: ...if contact with the element is made Setting the Required Water Temperature The water heating control circuit will retain the last used temperature set point set on the control panel so for many appli...

Страница 12: ...y glove when removing the lid as the lid may be hot enough to cause burns or other injury NEVER remove the lid with your face directly over the water bath Steam may be present and burns or injury coul...

Страница 13: ...placed below the level of the water bath on a firm hard flat surface If emptying into a sink or drain ensure you have clear and unobstructed access to it If the bath is fitted with a drain tap models...

Страница 14: ...ferring to the section in this User Manual labeled Filling The Water Tank and allow 1 minute for the bath to cool The water bath can now be re operated as per the section of this User Manual labeled S...

Страница 15: ...mple substances that may compromise the integrity or function of electrical insulation electrical components stainless steel components or water seals Ratek and the manufacturer are consulted prior to...

Страница 16: ...ous injury or risk to life Refer the equipment to an authorized service technician for repair Over Temperature Safety Cutout Test The over temperature safety cutout should be tested periodically to en...

Страница 17: ...to the up position marked with a circle and refill the water tank with 1 000mL of cold water to cool the element by referring to the section in this User Manual labeled Filling The Water Tank This wil...

Страница 18: ...to 2 000 Metres Temperature 5 Celsius to 40 Celsius Ambient temperature will limit the minimum achievable water temperature Maximum relative humidity 80 for temperatures up to 31 Celsius decreasing l...

Страница 19: ...ng The Water Bath provides a simple to operate user interface when used in conjunction with this User Manual If at any stage you experience abnormal operation anything other than that described in thi...

Страница 20: ...n approved repairer or Ratek Instruments at the discretion of Ratek Instruments The limit of liability shall extend to the repair of the product only all other compensation claims are excluded from th...

Страница 21: ...he unit to ensure safety for the service technicians Pack the unit into it s original packaging with the supplied mains power lead and use all original protective inserts If the original packaging is...

Отзывы: