EA4240-CL and EA4710-CL / USER MANUAL Rev1.3
February 2018
12
To adjust the image to upright, go to Resolution tab on the left and select left B-T from the drop down
menu. Then click on buffer tab again.
Figure 8b: Adjusted Live Image Capture Window
Changing the settings on XCAP
You have the ability to change several settings to control the camera through the tabs in the control
window of the EPIX software.
4.4
TEC
Thermoelectric Cooling (TEC). Use this tab to set and read back the sensor temperature. To operate
the camera for long exposures you will need to adjust the temperature of the Eagle V to deep cooling
using a combination of the TEC and a water re-circulator. The camera is capable of cooling to -90C.
Ensure that the camera is connected to the circulator and that the temperature is set to 20°C. Make
sure to rock the camera to ensure that the coolant is flowing evenly through the camera with no air
bubbles.
Then click on the TEC tab. On the TE Cooler drop down, select Enabled, and then set the TEC to -90°C.
This will take several minutes for the camera to reach the temperature. Click on the update temp button
every few minutes to check on the sensor temperature.