background image

 

14 

 

Sectioning procedures 

 After the specimen orientation and the blade angle have been set, trimming can be carried 

out using the trimming function. 

 Place the anti-roll plate against the blade. 

 Turn the hand wheel in a clockwise direction to carry out sectioning. 

  The tissue sections move into the space between the blade and the anti-roll plate. 

  Use the handle to move the anti-roll plate assembly away from the specimen. 

  The section should remain on the clamping plate. 

  Arrange specimen section gently with a brush. 

  Transfer the specimen onto a glass microscope slide with help of adhesion forces. 

Tips for Successful Sectioning 

 

Check the condition of the blade edge: move the blade horizontally to the left or right side to 

obtain a sharp cutting edge. 

 

Check the angle of anti-roll plate and correct it if necessary. 

 

Ensure that the edge of the anti-roll plate glass is clean and free of debris. 

 

Carefully remove frost from the front and the rear part of the blade, the anti-roll plate, and the 

clamping plate (i.e., with ethanol). 

 

Tighten all clamping screws and clamping levers on the blade holder and specimen head. 

 

Select the appropriate knife carrier and specimen temperature according to the specimen type 

o

 

See recommended Temperature Ranges for Sectioning of Different Types of Fresh 

Specimen on the next page. 

 

Allow time for the temperatures within the chamber to stabilize. 

 

Select the appropriate freezing compound. Avoid excessive use of freezing compound. 

o

 

Note:

 

If  the  specimen  was  frozen  with  liquid  nitrogen  or  similar  freezing  techniques,  the 

specimen must be allowed to adjust to the cutting temperature.

 

 

 

During  defrosting,  remove  the  specimens  from  the  cryochamber,  as  the  temperature  inside 

the  chamber  will  increase.  Do  not  leave  or  store  specimens inside  the  cryostat  over  a  long 

period of time. due to a power failure or other unexpected malfunctions of the instrument, the 

specimen might be damaged. 

 

Adjustment of proper blade angle. Select a blade angle of 8-12°. The preset clearance angle 

is 10°. 

 

Select an appropriate cutting speed: The harder the material, the slower the cutting speed. 

 

Be very careful when bringing the blade and specimen together. 

 

 

Содержание Basics CRY85

Страница 1: ...ttention Before using the instrument please read this Instruction carefully and keep the Instruction for further reference Please refer to the nameplate on the back of the instrument for the serial nu...

Страница 2: ...t 6 Middle Peltier freezing 7 Middle Standby 7 Middle Sectioning trimming setting 7 Right Clock setting 8 Right Temperature setting 8 Right Fast freezing 8 Right Start of work timer 9 Right End of wor...

Страница 3: ...oning procedures 14 Tips for Successful Sectioning 14 Recommended Temperature Ranges 15 5 Troubleshooting 16 6 Cleaning and Maintenance 17 Cleaning the display screens 17 Cleaning procedures 17 Empty...

Страница 4: ...men head cooling UV option enables surface disinfection of harmful bacteria viruses and spores by using 35 min after each use The Peltier semiconductor cooling function on the quick freeze shelf can b...

Страница 5: ...50 m Specimen retraction Adjustable range 0 80 m Increment 5 m Horizontal stroke 20 mm Vertical stroke 60 mm Voltage AC 110V 10 AC 220V 10 Frequency 50 60Hz Power 600W Largest start up current 5 secon...

Страница 6: ...s as well as the back of the instrument to facilitate ventilation and heat dispersion The instrument can be used after it is kept standing still for at least 2 hours Please ensure the grounding wire o...

Страница 7: ...e To protect the instrument from being damaged by an abrupt startup the instrument cannot be re started within 3 minutes after it is turned off The central operation panel includes three zones left mi...

Страница 8: ...utes which will count down to 14 13 2 1 until the Peltier stops cooling and the indicator light is off User may also press this button to stop the Peltier when it is running The Peltier stops running...

Страница 9: ...actual temp Peltier on off Clock setting The clock can be set by pressing D to adjust weekday and pressing H and M to adjust hour and minute respectively Temperature setting Press the temperature but...

Страница 10: ...rking day then press the H and M buttons to adjust hour and minute respectively The system will automatically set 24 hours of the standby mode for non working days Programming end of work timer Press...

Страница 11: ...le by moving forward Press button specimen is moving towards user slowly Fast Advance and Retract Press double arrow buttons to move the specimen head forward or backward The indicator light on the bu...

Страница 12: ...lly move the holder After the holder is placed in the desired position lock the lever clockwise Note Please lock the blade angle in a proper position before adjusting forward backward movements Blade...

Страница 13: ...d in place by two screws Ensure that the gap clearance side of the anti roll plate is facing inwards Position the anti roll plate in the holder gently pushing the glass to the bottom of the holder The...

Страница 14: ...eel Specimen Setup The specimen must be correctly secured and oriented in the specimen head for subsequent sectioning To secure and orient the specimen Loosen the specimen chuck fixation screw and ins...

Страница 15: ...late and the clamping plate i e with ethanol Tighten all clamping screws and clamping levers on the blade holder and specimen head Select the appropriate knife carrier and specimen temperature accordi...

Страница 16: ...t 35 C to 40 C 30 C or below Cervix 25 C to 30 C 20 C Connective Tissue 25 C 16 C Gut 25 C 20 C Heart 25 C 20 C Kidney 25 C 15 C Lip 25 C 13 C Liver 25 C 13 C Lung 25 C 15 C Lymph Node 25 C 13 C Muscl...

Страница 17: ...holder not correctly clamped Check tighten blade holder 2 Blade incorrectly clamped Blade clamping force can be increased by tightening clamp screw at the rear of the top stage 3 Specimen incorrectly...

Страница 18: ...lade holder Carefully clean and dry the dismounted blade holder As there is condensate humidity inside the microtome structure dry the microtome components inside the cryochamber very carefully For th...

Страница 19: ...ion of the instrument If the original packaging is no longer available please contact customer service via support rankinbiomed com II The conditions for storage and transportation must be observed du...

Страница 20: ...this instrument is designed to accepted standards of safety Its use does not entail any hazard if operated in accordance with the instructions given in the documentation However the following safety p...

Страница 21: ...Item Description Qty Special Note 1 Base Unit 1 2 Disposable Blade Holder 1 2 in 1 low and high profile 3 Handwheel Handle 1 4 Specimen Disc 3 5 Power Cord 1 6 Brush 2 7 Heated Glass Door 1 Removable...

Страница 22: ...ers dealers or resellers Limitation of Warranty The warranty stated above is the only warranty applicable to this product All other warranties expressor implied including all implied warranties of mer...

Отзывы: