Changing blade
Release the blade clamping lever on the upper right to open the blade clearance, insert a new
blade from left to right and then lock the clamp lever.
Adjust the angle of the blade
Release the angle adjustment level, adjust the blade holder to the desired angle and then tighten
the shaft.
Fit and adjust the anti-roll plate
The anti-roll Plexiglas plate is fitted following the below procedures:
•
The anti-roll plate is held in place by two screws.
•
Ensure that the gap clearance side of the anti-roll plate is
facing inwards.
•
Position the anti-roll plate in the holder, gently pushing the
glass to the bottom of the holder.
•
The edge of the anti-roll plate should be parallel to the edge of
the blade.
•
Tighten the screws.
Note: Microtome blades are extremely sharp and can cause severe
Injuries - please always use good laboratory practice when handling
them!
Note:
The cutting angle shall be no more than 10°
The edge of the anti-roll plate and edge of the blade must be flush.
10°
30°
Tissue
Adjustable angle
Angle of cutting blade edge
Anti rolling plate
Disposable blade