Ranger Irrigation Fluid Warming System
English
Slot for
warming cassette
Alarm
indicator
light
Power
switch
Alphanumeric
display
I.V. pole clamp
Product Description
The Ranger Irrigation Fluid Warming System consists of a warming unit and a sterile disposable fluid
delivery set.
RANGER Irrigation Fluid
Warming Unit
The warming unit is a compact, lightweight, liquid-resistant device with a clamp located on the
side for attachment to an I.V. pole. When mounted to the I.V. pole, the unit fits easily above the Bair
Hugger® 500 or 700 series warming unit. A carrying handle on the top of the unit makes it easy to
transport.
Located on the front panel you will find:
• Alphanumeric display that indicates the heater temperature during normal operation. In an over
temperature condition, the display alternately flashes a temperature of 48°C or higher and the word “HI.”
In an under-temperature condition, the display alternately flashes a temperature of 33˚C or lower and
word “LO.”
• Alarm indicator light that comes on when either an over- or under-temperature condition occurs.
RANGER Irrigation Fluid Warming Disposable Set
Irrigation fluid warming disposable sets
1
include a warming cassette (heat exchanger), tubing, connectors,
flow chamber and clamps.
1. Refer to the instructions provided with each disposable set for information on use.
2. With the bag hung 100 cm above the scope under gravity flow.
3. Includes patient line.
Model 27
Flow Rate
PRIMIng VoluMe
PatIent lIne length
Irrigation Set
0-865 mL/min
2
308 mL
3
213 cm
Model 27
Flow Rate
PRIMIng VoluMe
PatIent lIne length
Irrigation Set
0-865 mL/min
2
308 mL
3
213 cm
Fluid warming
cassette
White clamp
Outlet tube
Exit
Blue clamp
Inlet tube
Fluid source