Ranger Irrigation Fluid Warming System
English
11
Specifications
Physical Characteristics
warming unit
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long;
wt.: 7 lb. 7 oz. (3.4 kg)
classification
Classified under IEC 601-1 Guidelines (and other national versions of the Guidelines) as Class I, Type B,
Ordinary Equipment. Not suitable for use in the presence of flammable anesthetic mixtures with air or
with oxygen or nitrous oxide. Classified by Underwriters Laboratories Inc. with respect to electric shock,
fire, and mechanical hazards only, in accordance with UL 60601-1 and in accordance with Canadian/CSA
C22.2 No. 601.1.
Electrical characteristics
Temperature characteristics
input voltage
set point temperature
100-120 or 220-240 VAC
41°C
operating frequency
over-temperature alarm
100-120 VAC, 50/60 Hz
48°C
220-240 VAC, 50 Hz
maximum heating power
under-temperature alarm
900 W
33°C
fuse
over-temperature cutoff
10 Amp
50°C
Leakage Current
Meets leakage current requirements in accordance with UL 60601-1 and IEC 601-1.
Environmental Conditions
operating temperature range
15° to 40°C (59° to 104°F)
storage temperature range
-20° to 45°C (-4° to 113°F)
operating humidity
10 to 85% RH, noncondensing
atmospheric pressure range
50 kPa to 106 kPa