Building Blocks and RF Signal Flow
R&S
®
ZVAX-TRM
11
User Manual 1325.1268.02 ─ 04
R&S
ZVA/ZVT with two internal sources can perform bidirectional intermodulation
measurements without having to alter the RF connections.
The built-in monitor switch allows to use the combiner input from path 3 (path 4) as ref-
erence signal for port 3 (port 4), even if the combiner is looped into source paths 1 and
3 (source paths 2 and 4).
For the related options, see
Chapter 4.2.4, "Broadband Combiners"
2.4 Direct Generator/Receiver Access
In order to enable application-specific extensions, the R&S
ZVAX-TRM frontend pro-
vides access to the source and measurement paths of ports 1 and 2, similar to the
"Direct Access" connectors at the R&S
ZVA/ZVT. This allows to loop external compo-
nents (filters, attenuators, isolators, amplifiers etc.) into the related source and/or mea-
surement paths or to connect additional measurement instruments.
Direct generator/receiver access is already part of the base models (see
on page 27) and doesn't require any additional option.
The external RF path extensions can be bypassed using electronic transfer switches;
there is no need to jumper the respective IN/OUT connectors.
2.5 RF Signal Flow (w/o combiners)
If no combiner is switched into the signal path, ports 1 and 2 are physically provided by
the extension unit (ports 1 and 2 of the analyzer are not used):
RF Signal Flow (w/o combiners)