Command reference
R&S
®
ZNL/ZNLE
989
User Manual 1178.5966.02 ─ 19
LTSTamp
Timestamp in local time
FSMode
Returns the frequency sweep mode that was used during cali-
bration.
STEP:
Stepped mode (for all segments)
ANAL:
Swept mode (for at least one segment)
Because swept mode is limited to frequency sweeps this always
returns STEP.
Usage:
Query only
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Gets the number of available power calibrations (both source power and measurement
receiver).
Use
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
to retrieve the
related settings.
Suffix:
<Ch>
.
Number of the calibrated channel.
<PhyPt>
This suffix is ignored.
<Wv>
This suffix is ignored.
Usage:
Query only
Manual operation:
See
SOURce:POWer:CORRection:NREadings
<NoCalSweeps>
Sets a limit for the number of calibration sweeps in the source power flatness cal. The
command is valid for all channels and ports.
Equivalent command:
SOURce:POWer:CORRection:COLLect:AVERage[:COUNt]
Parameters:
<NoCalSweeps>
Number of readings
Range:
1 to 100
*RST:
2
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
"Flatness Cal – Max Iterations"
SOURce<Ch>:POWer:CORRection:OSOurces[:STATe]
<Boolean>
Switches off all other sources during the calibration sweep for channel
<Ch>
.
VNA command reference