background image

HI758

  

Marine Calcium

INSTRUCTION MANUAL

Reagent Sets

HI758-26

Reagents for 25 Marine Calcium tests

Other Accessories

HI758-11

Marine Calcium Certified Standard Kit

HI731315

Glass Cuvettes and Caps for Checker

®

 HC Colorimeters (2 pcs.)

HI731318

Cloth for wiping cuvettes (4 pcs.)

HI731339P

0.1 mL minipipette

HI731349P

Tip for 0.1 mL minipipette (10 pcs.)

HI740028P

1.5V AAA batteries (12 pcs.)

HI70436M

Deionized water (230 mL)

HI93703-50

Cuvette cleaning solution (230 mL)

Accessories

IST758   05/18

HI758

 is warranted for a period of one year after date of purchase against defects in 

workmanship and materials when used for their intended purpose and maintained according 
to instructions. This warranty is limited to repair or replacement free of charge. Damage due 
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service 
is required, contact your local Hanna Instruments Office. If under warranty, report the model 
number, date of purchase, serial number and the nature of the problem. If the repair is not 
covered by the warranty, you will be notified of the charges incurred. If the instrument is to 
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the 
Technical Service department and then send it with shipping costs prepaid. When shipping any 
instrument, make sure it is properly packed for complete protection.

Before using Hanna Instruments products, make sure that they are entirely suitable for 
your specific application and for the environment in which they are used. Operation of 
these instruments may cause unacceptable interferences to other electronic equipment, 
thus requiring the operator to take all necessary steps to correct such interferences. Any 
variation introduced by the user to the supplied equipment may degrade the instrument’s 
EMC performance. To avoid damages or burns, do not put the instrument in microwave 
oven. For yours and the instrument safety do not use or store the instrument in hazardous 
environments.

Recommendations for Users

Warranty

To conserve battery, the instrument shuts down after 
10 minutes of non-use and 2 minutes after reading. One fresh battery 
lasts for a minimum of 5000 measurements. When the battery is 
dead the instrument will display 

“bAd”

 then 

“bAt”

 for 1 second 

and then turns off. To replace the battery, turn the instrument off 
by holding the button until the meter shuts off. Turn the instrument 
upside down and remove the battery cover with a screwdriver. 
Remove the battery from its location and replace it with a new one, 
inserting the negative end first. Replace the battery cover and screw.

It is important that the sample does not contain any debris. 
Whenever the cuvette is placed into the measurement cell, it must be 
dry outside, and completely free of fingerprints, oil and dirt. Wipe it 
thoroughly with 

HI731318

 or a lint-free cloth prior to insertion.

Shaking the cuvette can generate bubbles, causing higher readings. 
To obtain accurate measurements, remove bubbles by swirling or by 
gently tapping the cuvette. Do not let the reacted sample stand for 
too long after reagent is added, as accuracy will be affected. After the 
reading, it is important to immediately discard the sample, otherwise 
the glass might become permanently stained. 

• Attach the pipette tip.
• Press the button down to the 

first stop

.

• Immerse the pipette tip in the liquid approximately 2-3 mm.
• Slowly let the button move back to the 

original position

, wait 2 seconds.

• Remove the pipette tip from the liquid.
• To dispense the liquid, place the pipette tip on the inside wall of the container.
• Slowly press the button down to the 

first stop

.

• Wait until all of the liquid has been dispensed.
• Press the button down to the 

second stop

, this will allow any remaining liquid to be dis-

pensed.

First stop Second stop

Original position

Minipipette Instructions:

Tips for an Accurate Measurement

Battery Management

Отзывы: