Functional Description

Errors and Warnings

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Under Range:

 A blinking “200” indicates that the reading is under 

range.

Over Range:

 A blinking value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Inverted Cuvette:

 The sample and the zero cuvette are inverted.

Low Battery:

 The battery is low and must be replaced soon. 

Dead Battery:

 The battery is dead and must be replaced. Once this 

indication is displayed, normal operation of the instrument will be 
interrupted. Change the battery and restart the meter.

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. 

Please check the preparation of the zero cuvette.

Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office. 
Each 

HI758

 meter is supplied complete with:

•  Sample Cuvettes and Caps (2 pcs.)
•  Reagents for 25 tests
•  1 mL syringe with tip 
•  Minipipette with tips
•  Plastic pipette
•  1.5V AAA battery
•  Instruction manual and Quick Reference Guide

Preliminary Examination

Range

200 to 600 ppm

Resolution

1 ppm

Accuracy

±6 ppm ±5% of reading @25 °C/77 °F

Light Source

Light Emitting Diode @610 nm

Light Detector

Silicon Photocell

Method

Adaptation of the zincon method.

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condesing

Battery Type

1.5V AAA (1 pc.)

Auto-Shut off

After 10 minutes of non-use

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

52 g (1.84 oz.)

Specifications

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction 
manual carefully before using the instrument.  
For more information about Hanna Instruments and our products, visit 
www.hannainst.com. 
For technical support, contact your local Hanna Instruments Office or e-mail us at 
[email protected]
Find your local Hanna Instruments Office on www.hannainst.com.

Measurement Procedure

Hanna Instruments reserves the right to modify the design, construction or appearance of 

its products without advance notice.

•  Turn the meter on by pressing the button. 

All segments will be displayed. When the 
display shows “

Add

”, “

C.1

” with “

Press

” 

blinking, the meter is ready.

•  Add 1 mL of 

HI7581

 Reagent A to the 

cuvette using the 1 mL syringe. Use 
the plastic pipette to fill the cuvette 
to the 10 mL mark with deionized 
water and replace the cap. Invert 
3-5 times to mix.

•  Place the cuvette into the meter and close 

the meter’s cap. Press the button. When the 
display shows 

“Add”

“C2”

 with 

“Press” 

blinking the meter is zeroed. 

•  Remove the cuvette from the meter and 

unscrew the cap. Use the minipipette to 
add 0.1 mL of sample to the cuvette.

•  Add 1 packet of 

HI7582

 Reagent B. 

Replace the cap and shake vigorously for 
15 seconds or until the powder is completely 
dissolved. Allow air bubbles to dissipate for 
15 seconds before taking a reading. 

•  Place the cuvette back into the meter and 

close the meter’s cap. Press the button. 
The meter displays concentration in ppm 
of calcium. The meter automatically turns 
off after 2 minutes.