Command reference
R&S
®
ZNL/ZNLE
976
User Manual 1178.5966.02 ─ 19
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
equivalent command
Manual operation:
See
"Flatness Cal – Max Iterations"
SOURce:POWer:CORRection:COLLect:AVERage:NTOLerance
<Tolerance>
Specifies the maximum deviation of the measured power from the target power of the
calibration. The command is valid for all channels and calibrated ports.
Parameters:
<Tolerance>
Tolerance value
Range:
0.001 dB to 1000 dB
*RST:
1 dB
Default unit: dB
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
SOURce:POWer:CORRection:COLLect:CFACtor
<Convergence>
Specifies the convergence factor for a source power calibration.
Parameters:
<Convergence>
Convergence factor
Range:
0 to 2
Increment: 0.1
Example:
See
SOURce<Ch>:POWer:CORRection[:ACQuire]
Manual operation:
See
SOURce:POWer:CORRection:COLLect:FLATness
<Boolean>
Enables or disables the source power calibration. The command is valid for all chan-
nels and ports.
It is not possible to disable flatness calibration and the reference receiver calibration
simultaneously (see
SOURce:POWer:CORRection:COLLect:RRECeiver
Parameters:
<arg0>
<Boolean>
Disables (OFF|0) or enables (ON|1) the flatness calibration. With
disabled flatness calibration, only one calibration sweep is per-
formed in order to calibrate the reference receiver; the previous
source calibration data is not overwritten.
*RST:
ON|1
VNA command reference