VNA GUI reference
R&S
®
ZNL/ZNLE
494
User Manual 1178.5966.02 ─ 19
Use "Detect Port Assignment" to detect the current port assignment (after reconnecting
the calibration unit).
Manual port assignment is not supported.
Validate
Runs the validation for the current port assignment. The validation proceeds through
the connected ports cal unit ports p
∈
{1, ..., 4} in ascending order, measuring S
pp
for
the Open, Short and Match standard.
●
"PASS": all measured reflection parameters are within the specified
, compared to the selected reference
●
"FAIL" otherwise.
[In this case, the validation stops at the first failed reflection measurement.]
Remote command:
CALCulate:CALValidate:RUN
CALCulate:CALValidate:RUN:RESult?
Advanced Mode
Switches to
.
Advanced mode
In advanced mode you can configure the validation settings (
) and run or rerun particular measurements (
,
Cal softtool