Command reference
R&S
®
ZNL/ZNLE
1035
User Manual 1178.5966.02 ─ 19
<SFId2>
Service function identifier entered as a (pseudo-)numeric value,
the dots being replaced by commas. Five groups of numbers are
allowed at maximum. See also
FORMat:DEXPort:SOURce
<Format>
Defines the format for traces retrieved with the R&S ZVR-compatible command
TRACe[:DATA][:RESPonse][:ALL]?
.
This command is not relevant for results read with the
CALCulate<Chn>:DATA...
commands.
Parameters:
<Format>
FDATa | SDATa | MDATa
See list of parameters below. The unit is the default unit of the
measured parameter; see
Range:
Depending on the measured parameter and format.
*RST:
SDATa
Example:
See
TRACe[:DATA][:RESPonse][:ALL]?
The following parameters are related to trace data:
FDATa
Formatted trace data, according to the selected trace format (
). 1 value per trace point for Cartesian diagrams, 2 values for polar dia-
grams.
SDATa
Unformatted trace data: Real and imaginary part of each measurement point. 2
values per trace point irrespective of the selected trace format. The trace mathe-
matics is not taken into account.
MDATa
Unformatted trace data (see
SDATa
) after evaluation of the trace mathematics.
INPut<PhyPt>:ATTenuation
<Attenuation>
Sets the attenuation for the received waves. This command is available if at least one
of the
Note
:
●
INPut<PhyPt>:ATTenuation
is not channel-specific; the value is valid for
all
channels
. Use
specific attenuation value.
Suffix:
<PhyPt>
.
Physical port number; if unspecified the numeric suffix is set to 1
VNA command reference