Command reference
R&S
®
ZNL/ZNLE
1016
User Manual 1178.5966.02 ─ 19
<SecondCal>
String parameter with the channel name (for <LocationSecond-
Cal>=CHANnel) or the calibration file name (for <LocationSe-
condCal>=POOL) of the first calibration.
<LocationSecondCal>CHANnel | POOL
Setting parameters:
<OutputCal>
String parameter with the channel name (for <LocationOutput-
Cal>=CHANnel) or the calibration file name (for <LocationOut-
putCal>=POOL) of the resulting calibration.
Note:
If (<OutputCal>,<LocationOutputCal>) matches one of the
input calibrations, this input calibration is replaced.
<LocationOutputCal> CHANnel | POOL
Return values:
<Joinable>
SYSTem:DATE
<Year>, <Month>, <Day>
The command queries or defines the instrument's current date setting.
Parameters:
<Year>
Year, four-digit number
<Month>
Month, two-digit number, 01 (for January) to 12 (for December)
<Day>
Day, two-digit number, 01 to the number of days in the month
Example:
SYST:DATE?
Response:
2012,05,01
- it is the 1st of May, 2012.
SYSTem:DISPlay:CONDuctances
<Boolean>
Changes the presentation of "capacitance C<i> in parallel with resistance R<i>" circuit
blocks in lumped de/embedding networks.
Parameters:
<Boolean>
ON - display conductances
OFF - display capacitaces
Manual operation:
See
"Conductance in Embedding Networks"
SYSTem:FREQuency?
<MinMax>
Queries the minimum and maximum frequency of the network analyzer. For an over-
view refer to the tables at the beginning of
Tip:
?
and the other sweep range
commands,
SYSTem:FREQuency?
can be used in all sweep modes.
Query parameters:
<MinMax>
MINimum | MAXimum
VNA command reference