Command reference
R&S
®
ZNL/ZNLE
941
User Manual 1178.5966.02 ─ 19
[SENSe<Ch>:]CORRection:STIMulus?
Queries the stimulus values of the active calibration. A calibration must be selected
before the command is executed; see example.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
*RST; :CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:STIM?
Query the stimulus frequencies. The response contains 201 fre-
quency values.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply it to the
active channel.
Usage:
Query only
Manual operation:
See
[SENSe<Chn>:]CORRection:SSTate?
Returns the system error correction state label of the active trace in channel
<Chn>
.
The response is a string variable containing the calibration state label in the trace list
('CAL', 'CAI', 'CAL OFF' ..; see
Chapter 9.5.4, "Calibration state labels"
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
See
Usage:
Query only
Manual operation:
See
[SENSe:]FREQuency...
The
[SENSe:]...
commands set frequency-related parameters, especially the mea-
surement and display ranges for the different sweep types. The frequency ranges for
the different instrument models are listed below; for more details refer to the data
sheet.
Table 11-18: Frequency ranges of R&S
ZNL analyzers
Frequency settings
Start, Stop
Center
Span
R&S
ZNL3
5 kHz to 3.0 GHz
>5 kHz to <3.0 GHz
1 Hz to 2.999995 GHz
R&S
ZNL4
5 kHz to 4.5 GHz
>5 kHz to <4.5 GHz
1 Hz to 4.499995 GHz
R&S
ZNL6
5 kHz to 6 GHz
>5 kHz to <6 GHz
1 Hz to 5.999995 GHz
VNA command reference