Command Reference
R&S
®
ZNC
693
User Manual 1173.9557.02 ─ 13
Parameters:
<CalType>
FRTRans | FTRans | RTRans | TOM | TSM | TRM | TRL | TNA |
TOSM | ETOM | ETSM | FOPort1 | FOPort2 | FOPort12 |
FOPTport | ROPTport | REFL1 | REFL2 | REFL12 | TPORt |
UOSM
Calibration types, TOM, TRM, TRL, TNA, TOSM, Full One Port,
One Path Two Port, Normalization (REFL1, REFL2 and REFL12
for one-port, FRTRans, FTRans, RTRans, and TPORT for two-
port), TOSM with unknown through.
The numbers in the parameters denote the analyzer ports. Param-
eters for two-port calibration types contain no numbers because
the command is only valid for ports 1 and 2.
Example:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE:DEFault
Generates a set of default system error correction data for the selected ports and cali-
bration type. The default data set corresponds to a test setup which does not introduce
any systematic errors; none of the measurement results acquired previously (
) is taken into account.
Tip:
The main purpose of the default correction data set is to provide a dummy system
error correction which you can replace with your own, external correction data. You may
have acquired the external data in a previous session or even on an other instrument. If
you want to use the external correction data on the analyzer, simply generate the default
data set corresponding to your port configuration and calibration type and overwrite the
default data. For details refer to the program example below.
Note:
This command must be used in combination with the R&S ZVR-compatible com-
and
if you want to use R&S ZNC-specific calibration commands or if
you want to calibrate more than 2 ports.
Suffix:
<Ch>
.
Channel number of the calibrated channel
R&S ZVR/ZVAB Compatible Commands