Command Reference
R&S
®
ZNC
692
User Manual 1173.9557.02 ─ 13
Setting parameters:
<Standard>
THRough | OPEN1 | OPEN2 | OPEN12 | SHORt1 | SHORt2 |
SHORt12 | MATCh1 | MATCh2 | MATCh12 | NET | ATT |
IMATch12 | REFL1 | REFL2 | SLIDe1 | SLIDe2 | SLIDe12 | LINe1 |
LINe2 | LINe3 | M1O2 | O1M2 | OSHort1 | OSHort11 | OSHort12 |
OSHort13 | OSHort2 | OSHort21 | OSHort22 | OSHort23 | M1S2 |
S1M2 | UTHRough
Standard types: Through (between port 1 and 2), Open, Short,
Match (MATCh12 and IMATch12 are synonymous), Symmetric
Network (NET), Attenuation (ATT), Reflect, Sliding Match (SLIDe),
Line1 (LINe1 and LINE are synonymous), Line2 and Line3 (esp.
for TRL calibration), Match/Open (M1O2, O1M2), Match/Short
(M1S2, S1M2), Offset Short (OSHort), Unknown Through
(UTHRough), ISOLation (optional, between port 1 and 2).
The numbers in the parameter names denote the analyzer ports.
Two numbers 12 mean that two separate calibrations are per-
formed at ports 1 and 2. For Offset Short standards, the first num-
ber denotes the port (1 or 2), the second number denotes the
number of the standard (1 to 3).
*RST:
ON
<Dispersion>
Optional status parameter for UTHRough standard:
OFF - unknown through standard is non-dispersive.
ON - unknown through standard is dispersive.
*RST:
OFF
<Delay>
Optional entry of delay time or phase for UTHRough standard:
<numeric> - entry of the delay time in ps (for non-dispersive stand-
ards) or of an estimate of the phase at the start frequency of the
sweep in deg (for dispersive standards). See also background
information for
AUTO - the analyzer determines the delay time or phase during
the calibration sweep.
*RST:
AUTO
Example:
See
Usage:
Setting only
[SENSe<Ch>:]CORRection:COLLect:METHod
<CalType>
Selects a one-port or two-port calibration type for channel
<Ch>
at ports 1/2.
Tip:
to select the calibration type for arbitrary analyzer ports or a multiport
calibration type.
Suffix:
<Ch>
.
Channel number of the calibrated channel.
R&S ZVR/ZVAB Compatible Commands