Command Reference
R&S
®
ZNC
634
User Manual 1173.9557.02 ─ 13
SOURce<Ch>:POWer<PhyPt>:CORRection:PPOWer
<RecCalPower>
Defines the source power which the R&S ZNC uses to perform the first calibration sweep
of the source power calibration ("Reference Receiver Cal Power"). The power value is
ignored if the R&S ZNC is set to use the port power result (
CPOW
).
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the param-
eter is valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the parameter
is valid for all sources.
Parameters:
<RecCalPower>
Reference receiver cal power
Range:
Depending on source power range of the analyzer
and the power range of the external power meter.
*RST:
0 dBm
Default unit: dBm
Example:
See
.
Manual operation:
See
"Reference Receiver Cal Power"
SOURce<Ch>:POWer<PhyPt>:CORRection:PSELect
<CalPower>
Qualifies how to define the source power which the R&S ZNC uses to perform the first
calibration sweep of the source power calibration ("Reference Receiver Cal Power").
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the param-
eter is valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the parameter
is valid for all sources.
Parameters:
<CalPower>
CPOWer | PPOWer
PPOWer
–
use cal power value defined via
CPOWer
–
use port power value
*RST:
PPOWer
Example:
See
.
Manual operation:
See
"Reference Receiver Cal Power"
SCPI Command Reference