Command Reference
R&S
®
ZNC
493
User Manual 1173.9557.02 ─ 13
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:ZOOM:BOTTom
<LowEdge>[,
<TraceName>]
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:ZOOM:TOP
<UppEdge>[,
<TraceName>]
These commands specify the lower and upper edge of the zoom window. In contrast to
manual control, all or part of the zoom window may be outside the original diagram.
Suffix:
<Wnd>
.
Number of an existing diagram (defined by means of
ON
). This suffix is ignored if
the optional
<TraceName>
parameter is used.
<WndTr>
Existing trace number, assigned by means of
the optional
<TraceName>
parameter is used.
Parameters:
<UppEdge>
Lower or upper edge of the zoom window.
Range:
<LowEdge>: -390 dB to +9.99 dB; <UppEdge>:
-89.99 dB to +232 dB
*RST:
<LowEdge>: -90 dB; <UppEdge>: +10 dB
Default unit: dB
<TraceName>
Optional string parameter for the trace name, e.g. 'Trc4'. If this
optional parameter is present, both numeric suffixes are ignored
(trace names must be unique across different channels and win-
dows).
Example:
See
Manual operation:
See
6.3.6 FORMat Commands
The
FORMAT...
commands select a data format for transferring numeric data (including
arrays) from and to the analyzer.
FORMat:BORDer
<ByteOrder>
Controls whether binary data is transferred in normal or swapped byte order.
SCPI Command Reference