Setting parameters:
<UnitRef>
COMPlex | MAGNitude | PHASe | REAL | IMAGinary | SWR |
GDELay | L | C
Keyword for the physical unit of the response values; dimension-
less numerss, relative power, phase, time, inductance, capaci-
tance units.
*RST:
n/a
Default unit: 1 (U, for COMPlex, REAL, IMAGinary, and SWR); dB
(for MAGNitude), deg (for PHASe), s (for GDELay),
H (Henry, for L), F (Farad, for C).
Example:
*RST; CALC:RIPP:DATA 1, 1500000000, 2000000000,
3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:RDOM:FORM COMP
Delete the ripple limit range, select complex units for the ripple
limit.
Usage:
Setting only
CALCulate<Chn>:RIPPle:SEGMent:COUNt?
Queries the number of ripple limit ranges. The response is an integer number.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Example:
*RST; CALC:RIPP:DATA 1, 1500000000, 2000000000,
3, 1, 2000000000, 3000000000, 5
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
Define and enable a second ripple limit range in the stimulus range
between 2 GHz and 3 GHz, assigning a ripple limit of +5 dB.
CALC:RIPP:SEGM:COUNT?
Query the number of ranges. The response is
2
.
Usage:
Query only
Manual operation:
See
CALCulate<Chn>:RIPPle:SEGMent<Seg>:LIMit
<Limit>
Defines the ripple limit for ripple limit range no.
<Seg>
. A range must be created first to
"Rules for Ripple Test Definition"
Tip:
To define several ripple limit ranges with a single command, use
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.