GUI Reference
R&S
®
ZNC
202
User Manual 1173.9557.02 ─ 13
Some of the "Target Search" functions are equal to the corresponding "Marker Search"
functions. Refer to the following sections:
●
chapter 4.2.6.4, "Search Range (Dialog)"
●
●
Target Value
Specifies the target value for the search.
Since V1.80 of the VNA software the target value can be specified in different formats
(see
below). For instance, this enables searching for particular phase
values in Smith charts.
Remote command:
Target Format
Selects the format in which the
shall be specified.
The selected target format applies to the current marker only: each marker may have a
different target format. The table below gives an overview on how a complex target value
z = x + jy is converted.
Target Format
Description
Formula
Lin Mag
Magnitude of z, unconverted.
|z| = sqrt ( x
2
+ y
2
)
dB Mag [dB]
Magnitude of z in dB
Mag(z) = 20 log|z| dB
Phase [°]
Phase of z
φ (z) = arctan (y/x)
Phase unwrap [°]
Unwrapped phase of z comprising
the complete number of 360°
phase rotations
Ф(z) = φ (z) + 2k·360°
Real
Real part of z
Re(z) = x
Imag
Imaginary part of z
Im(z) = y
Trace Settings