Command Reference
R&S
®
ZNB/ZNBT
805
User Manual 1173.9163.02 ─ 55
Parameters:
<Boolean>
*RST:
OFF
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
CALCulate<Chn>:EYE:JITTer:TYPE:PERiodic
<Boolean>
Enables/disables periodic jitter insertion in the generator simulation of the related eye
diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
Amplitude, frequency and phase of the jitter can be specified using
CALCulate<Chn>:EYE:JITTer:PERiodic:MAGNitude
,
and
CALCulate<Chn>:EYE:JITTer:PERiodic:
, respectively.
Note that jitter will not be applied unless the jitter functionality is activated using
CALCulate<Chn>:EYE:JITTer:STATe
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
*RST:
OFF
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
CALCulate<Chn>:EYE:JITTer:TYPE:RANDom
<Boolean>
Enables/disables random jitter insertion in the generator simulation of the related eye
diagram.
This command will raise an execution error if the active trace in the selected channel is
not an eye diagram.
The standard deviation can be set using
CALCulate<Chn>:EYE:JITTer:RANDom:
.
Note that jitter will not be applied unless the jitter functionality is activated using
CALCulate<Chn>:EYE:JITTer:STATe
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
*RST:
OFF
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
SCPI Command Reference