Function of the Status LED

R&S

®

ZN-Z103

6

User Manual 1321.2276.02 ─ 02

Excessive Input Power

If you use an external power amplifier, make sure that the maximum RF
input power of the calibration unit quoted in the data sheet is never excee-
ded.

4

Function of the Status LED

The LED on top of the calibration unit informs about the actual status of the
device. The different states have the following meaning:

OFF

The calibration unit is not connected or defective.

Red

The microcontroller is running but there is no USB communication with the

network analyzer. With an actual network analyzer firmware this status should

only appear for a short time after connecting the calibration unit.

Green

The calibration unit is ready use.

Blinking orange Data transfer between the calibration unit and the network analyzer. Do not

disconnect the USB cable.

Blinking blue

Calibration in progress.

5

Performing an Automatic Calibration

The analyzer uses a reflection calibration type (e.g. Full One Port / Refl OSM).
For details refer to your analyzer’s operating manual or help system.

After establishing an 

USB Connection to the Analyzer

 and initialization, you can

use the calibration unit as follows:

1. Connect an RF port of the analyzer to the RF port of the calibration unit.

2. Perform the reflection calibration of the selected port using the automatic cali-

bration function of the analyzer. For details refer to the analyzer’s operating
manual or help system.

 

 

Содержание ZN-Z103

Страница 1: ...R S ZN Z103 Calibration Unit User Manual User Manual 1321 2276 02 02 EF 2 Test Measurement...

Страница 2: ...bH Co KG M hldorfstr 15 81671 M nchen Germany Phone 49 89 41 29 0 Fax 49 89 41 29 12 164 Email info rohde schwarz com Internet www rohde schwarz com Subject to change Data without tolerance limits is...

Страница 3: ...are electronically switched when a calibration is performed The calibration kit data for the internal standards is stored in the calibration unit so that the analyzer can calculate the error terms an...

Страница 4: ...ty considerations To maintain this condition and to secure safe operation you must observe all instructions and warnings given in this manual and in the user manual of the Rohde Schwarz network analyz...

Страница 5: ...the connecting USB cable should not exceed 3 m It is recommended to use the cable delivered with the calibration unit You can connect cal units and other devices mouse USB memory stick etc simultaneo...

Страница 6: ...y appear for a short time after connecting the calibration unit Green The calibration unit is ready use Blinking orange Data transfer between the calibration unit and the network analyzer Do not disco...

Страница 7: ...observe the following items No adaptors must be inserted between the calibration unit and the test port After connecting the unit to the USB port allow for a sufficient warm up time see data sheet be...

Отзывы: