R&S ZN-Z103 Скачать руководство пользователя страница 5

Connecting the Calibration Unit

R&S

®

ZN-Z103

5

User Manual 1321.2276.02 ─ 02

A single RF connector type N (m).

USB Connection to the Analyzer

Follow the steps below and the instructions given in the analyzer’s operating
manual.

1. Switch on and start up your network analyzer.

2. Connect the USB type A connector of the USB cable to any of the USB type A

connectors of the network analyzer. You can also connect the unit before
switching on the analyzer.

3. Wait until the operating system has recognized and initialized the new hard-

ware. After the initialization is completed, the status LED will turn green.

The unit is now ready to be used in an automatic calibration, as outlined
below.

The length of the connecting USB cable should not exceed 3 m. It is
recommended to use the cable delivered with the calibration unit.

You can connect cal units and other devices (mouse, USB memory stick
etc.) simultaneously.

An unused calibration unit may remain connected to the USB port while
the network analyzer is performing measurements. It must be discon-
nected during a firmware update.

It is safe to connect or disconnect the calibration unit while the network
analyzer is operating. Never connect or disconnect the unit while data is
being transferred between the analyzer and the unit. Never connect the
unit during a firmware update.

RF Connection

See 

Chapter 5, "Performing an Automatic Calibration"

on page 6.

The maximum RF input power of the calibration unit is beyond the RF output
power range of a network analyzer, so there is no risk of damage if the device is
directly connected to the test ports.

Содержание ZN-Z103

Страница 1: ...R S ZN Z103 Calibration Unit User Manual User Manual 1321 2276 02 02 EF 2 Test Measurement...

Страница 2: ...bH Co KG M hldorfstr 15 81671 M nchen Germany Phone 49 89 41 29 0 Fax 49 89 41 29 12 164 Email info rohde schwarz com Internet www rohde schwarz com Subject to change Data without tolerance limits is...

Страница 3: ...are electronically switched when a calibration is performed The calibration kit data for the internal standards is stored in the calibration unit so that the analyzer can calculate the error terms an...

Страница 4: ...ty considerations To maintain this condition and to secure safe operation you must observe all instructions and warnings given in this manual and in the user manual of the Rohde Schwarz network analyz...

Страница 5: ...the connecting USB cable should not exceed 3 m It is recommended to use the cable delivered with the calibration unit You can connect cal units and other devices mouse USB memory stick etc simultaneo...

Страница 6: ...y appear for a short time after connecting the calibration unit Green The calibration unit is ready use Blinking orange Data transfer between the calibration unit and the network analyzer Do not disco...

Страница 7: ...observe the following items No adaptors must be inserted between the calibration unit and the test port After connecting the unit to the USB port allow for a sufficient warm up time see data sheet be...

Отзывы: