Preparing for use
R&S
®
SZM
20
Manual 1179.3967.02 ─ 02
► Connect the USB cable to the "USB" connector of the R&S
SZM.
5.8
Connecting to RF In
The "RF In" connector is a 2.92
mm female connector. The connector is located
on the rear panel of the R&S
SZM.
Use an RF cable that matches the connector. See
To connect the RF cable with the connector
1.
NOTICE!
Excessive reverse power at the RF connector can damage the
instrument.
Make sure that the signal power is within the limits as given in the data sheet.
2.
NOTICE!
DC voltage at the RF connector can damage the instrument. Never
apply DC voltage to the RF input connectors.
Make sure that the values are within the DC limits given in the data sheet.
3. If your test setup has a DC component at the RF input, insert a DC blocker.
4. Align the connector of the cable with the "RF In" connector of the R&S
SZM.
5. Carefully join the connectors so the pin of the device connector fits in the
socket of the cable connector.
6.
NOTICE!
Excessive tightening can damage the connectors.
Torque the nut to the specified limit using a calibrated torque wrench. Hold the
opposite connector part stationary with a spanner.
For applicable torque values, see the "Handling" chapter of the application
note 1MA99:
https://www.rohde-schwarz.com/appnote/1MA99
5.9
Connecting to the DUT
The R&S
SZM can be equipped with an optional test port adapter for the suppor-
ted frequency band. The test port adapter is the interface to the DUT that pro-
vides the output of the multiplied RF frequency.
Connecting to the DUT